A long-short double pulse laser-induced breakdown spectroscopy (long-short DP-LIBS) method was employed to improve the analytical performance of LIBS for the measurement of manganese in steel samples. The long pulse with a duration of 60 μs was generated using a neodymium-doped yttrium aluminum garnet (Nd:YAG) laser which was operated at free-running (FR) mode. To investigate the detection ability without sample preparation, the steel washers were tested using single-pulse LIBS (SP-LIBS) and long-short DP-LIBS, respectively. The measurement results show that long-short DP-LIBS was able to record clear spectra from the steel washers with a surface layer. Through the observation on the laser craters with a scanning electron microscope (SEM), the results suggest that the improvement in detection ability can be attributed to the pre-irradiation effect of long-pulse laser beam. Next, the analytical performance for quantitative measurement of manganese was evaluated employing ten standard steel samples. The results show that the linearity fit (R) of the calibration curve is 0.988 for long-short DP-LIBS, whereas, R is only 0.810 for SP-LIBS under the same measurement conditions. The repeated measurement results show that the average relative standard deviation (RSD) of the tested samples is 29.3% for SP-LIBS and is 10.5% for long-short DP-LIBS. The prediction results also show that the average relative error of prediction (REP) is 94.9% for SP-LIBS and is 4.9% for long-short DP-LIBS.

Download full-text PDF

Source
http://dx.doi.org/10.1177/0003702818803943DOI Listing

Publication Analysis

Top Keywords

long-short dp-libs
24
steel samples
12
manganese steel
8
long-short
8
long-short double
8
double pulse
8
pulse laser-induced
8
laser-induced breakdown
8
breakdown spectroscopy
8
analytical performance
8

Similar Publications

The long-short double-pulse laser-induced breakdown spectroscopy (LS-DP-LIBS) method was applied to qualitative and quantitative analyses of underwater steel samples to improve the detection ability of underwater measurement. The stable plasma intensity and discrete emission lines were detected using LS-DP-LIBS when comparing the measured results of single-pulse LIBS (SP-LIBS) and LS-DP-LIBS. The long pulse produces a cavitation bubble without plasma, and the short pulse induces the plasma of steel samples within the bubble.

View Article and Find Full Text PDF

A long-short double pulse laser-induced breakdown spectroscopy (long-short DP-LIBS) method was employed to improve the analytical performance of LIBS for the measurement of manganese in steel samples. The long pulse with a duration of 60 μs was generated using a neodymium-doped yttrium aluminum garnet (Nd:YAG) laser which was operated at free-running (FR) mode. To investigate the detection ability without sample preparation, the steel washers were tested using single-pulse LIBS (SP-LIBS) and long-short DP-LIBS, respectively.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!