Objectives: To evaluate differences in radiation dose and image quality across institutions, fluoroscope vendors and generations of fluoroscopes for pediatric cardiac catheterization.

Background: Increased recognition of the potentially harmful effects of ionizing radiation has spurred technological advances in fluoroscopes, as well as increased focus on optimizing fluoroscope performance. There is currently little understanding of variability in the dose-image quality relationship across institutions, fluoroscope vendor and/or generation of equipment.

Methods: We evaluated latest generation fluoroscopes from Phillips, Siemens, GE, and Toshiba, and an older generation Phillips fluoroscope (release date 2003) at three different institutions. Radiation dose was measured using an anthropomorphic dose-assessment phantom with effective dose in mSv estimated from Monte Carlo simulations. Image quality phantom images were scored on a 12-point scale by three blinded reviewers.

Results: Fluoroscope effective doses ranged from 0.04 to 0.14 mSv/1,000 pulses for fluoroscopy with associated composite image quality scores ranging from 8.0 ± 0.6 to 10.4 ± 1.3. For cineangiography, effective doses ranged from 0.17 to 0.57 mSv/1,000 frames with image quality scores ranging from 10.1 ± 0.3 to 11.1 ± 0.3. There was modest correlation between effective dose and image quality (r = 0.67, P = 0.006). The older generation fluoroscope delivered consistently higher doses than the newer generation systems (2.3- to 3.5-fold higher for fluoroscopy; 1.1- to 3.4-fold higher for cineangiography) without appreciable differences in image quality.

Conclusion: Technological advances have markedly improved fluoroscope performance. Comparing latest generation systems across vendors and institutions, we found variability in the dose-IQ relationship and speculate that this reflects both equipment and institutional optimization practices.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6294676PMC
http://dx.doi.org/10.1002/ccd.27793DOI Listing

Publication Analysis

Top Keywords

image quality
24
radiation dose
12
dose image
12
vendors generations
8
institutions fluoroscope
8
technological advances
8
fluoroscope performance
8
latest generation
8
older generation
8
effective dose
8

Similar Publications

Purpose: Our study evaluated skeletal muscle mass, function and quality among mild autonomous cortisol secretion (MACS) patients and non-functioning adrenal incidentaloma (NFAI) patients in comparison with the control group without adrenal mass.

Methods: 63 NFAI (49 female, 14 male) and 31 MACS (24 female, 7 male) patients were included in the study. As the control group, 44 patients (31 women, 13 men) who were known to have no radiological adrenal pathology on computed tomography or magnetic resonance imaging performed for other reasons were selected.

View Article and Find Full Text PDF

18F-Sodium Fluoride PET/CT as a Tool to Assess Enthesopathies in X-Linked Hypophosphatemia.

Calcif Tissue Int

January 2025

Endocrinology Department, School of Medicine, Pontificia Universidad Católica de Chile, Av. Diagonal Paraguay 262, Cuarto Piso, Santiago, Chile.

X-linked hypophosphatemia (XLH) is a rare metabolic disorder characterized by elevated FGF23 and chronic hypophosphatemia, leading to impaired skeletal mineralization and enthesopathies that are associated with pain, stiffness, and diminished quality of life. The natural history of enthesopathies in XLH remains poorly defined, partly due to absence of a sensitive quantitative tool for assessment and monitoring. This study investigates the utility of 18F-NaF PET/CT scans in characterizing enthesopathies in XLH subjects.

View Article and Find Full Text PDF

Correlations between spinopelvic parameters and health-related quality of life in degenerative lumbar scoliosis patients before and after long -level fusion surgery.

BMC Musculoskelet Disord

January 2025

Department of Orthopedics, Peking University Third Hospital, No. 49. North Garden Street, Hai Dian District, Beijing, 100191, People's Republic of China.

Background: For degenerative lumbar scoliosis (DLS), prior studies mainly focused on the preoperative relationship between spinopelvic parameters and health-related quality of life (HRQoL), lacking an exhaustive evaluation of the postoperative situation. Therefore, the postoperative parameters most closely bonded with clinical outcomes has not yet been well-defined in DLS patients. The objective of this study was to comprehensively assess the correlation between radiographic parameters and HRQoL before and after surgery, and to identified the most valuable spinopelvic parameters for postoperative curative effect.

View Article and Find Full Text PDF

Although the Transformer architecture has established itself as the industry standard for jobs involving natural language processing, it still has few uses in computer vision. In vision, attention is used in conjunction with convolutional networks or to replace individual convolutional network elements while preserving the overall network design. Differences between the two domains, such as significant variations in the scale of visual things and the higher granularity of pixels in images compared to words in the text, make it difficult to transfer Transformer from language to vision.

View Article and Find Full Text PDF

Functional magnetic resonance imaging (fMRI) has dramatically advanced non-invasive human brain mapping and decoding. Functional near-infrared spectroscopy (fNIRS) and high-density diffuse optical tomography (HD-DOT) non-invasively measure blood oxygen fluctuations related to brain activity, like fMRI, at the brain surface, using more-lightweight equipment that circumvents ergonomic and logistical limitations of fMRI. HD-DOT grids have smaller inter-optode spacing (~ 13 mm) than sparse fNIRS (~ 30 mm) and therefore provide higher image quality, with spatial resolution ~ 1/2 that of fMRI, when using the several source-detector distances (13-40 mm) afforded by the HD-DOT grid.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!