Background: The associations between mechanical circulatory support (MCS), acquired von Willebrand syndrome (AvWS), and clinical outcome are incompletely understood.

Methods: In 128 heart failure patients with pulsatile MCS implants (65 total artificial heart or biventricular assist device implants, 63 left ventricular assist device [LVAD] implants) and 76 patients with continuous flow LVAD implants, we analyzed the von Willebrand factor (vWF) profile before (≤24 h) and 17.5 (standard deviation: 5.1) days after device implant. We determined vWF concentrations, vWF activity, and vWF collagen binding capacity and calculated ratios of vWF activity/binding capacity with vWF concentration. The relation of the vWF profile with clinical outcomes such as stroke, gastrointestinal bleeding, and survival was also evaluated. Events were assessed up to 1 year of device implant.

Results: All entities of vWF were already significantly elevated preoperatively and remained high after MCS implantation. The ratios of vWF activity/concentration (vWF:RCo/Ag) and collagen binding capacity/concentration (vWF:CBA/Ag) were significantly reduced preoperatively and remained low postoperatively, indicating AvWS. The preoperative alterations in the vWF profile were already present in patients without intra-aortic balloon pump and/or extracorporeal circulatory membrane oxygenation implants. The vWF profile was unrelated to postoperative stroke. However, a higher postoperative ratio of vWF:CBA/Ag was independently associated with increased gastrointestinal bleeding. In addition, a postoperative increase in vWF concentrations and activity were independent predictors of increased 1-year mortality.

Conclusions: Our data indicate that AvWS is present in heart failure patients before device implantation, and is independently associated with clinical outcomes, especially with 1-year mortality.

Download full-text PDF

Source
http://dx.doi.org/10.1111/jocs.13794DOI Listing

Publication Analysis

Top Keywords

vwf profile
16
von willebrand
12
clinical outcomes
12
vwf
12
willebrand factor
8
mechanical circulatory
8
circulatory support
8
heart failure
8
failure patients
8
assist device
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!