Background: Optimal implantation results of bioresorbable scaffolds (BRS) are typically assumed to require postdilatation with non-compliant (NC) balloons to achieve full scaffold apposition and minimize event rates. We systematically evaluated the mechanical effect of NC balloon postdilatation on magnesium-based BRS (Magmaris®, Biotronik AG, Bülach, Switzerland) in vivo.

Methods: In 35 patients, 40 Magmaris® BRS were implanted to treat 37 de novo coronary artery stenoses. A systematic implantation protocol was followed. After appropriately sized NC balloon predilatation (1:1:1 vessel:balloon:scaffold ratio), Magmaris® BRS were implanted with a pressure of 10 atm, followed by NC balloon postdilatation at nominal BRS size (standardized at 16 atm). OCT was performed before and after postdilatation and OCT images were analyzed at a spacing of 0.2 mm to measure BRS dimensions and determine apposition as well as to detect strut fractures.

Results: PCI with Magmaris® BRS (mean diameter: 3.21 ± 0.32 mm; mean length: 20 ± 4 mm) was successful in all cases, in one case a non-flow-limiting distal edge dissection occurred after implantation and before postdilatation. NC balloon postdilatation led to significantly larger mean scaffold diameter (3.21 ± 0.32 mm vs. 2.80 ± 0.39 mm, P < 0.001), abluminal scaffold area (7.92 ± 1.43 mm vs. 6.72 ± 1.28 mm , P < 0.001) and lumen area (7.58 ± 1.1 mm vs. 6.83 ± 1.12 mm , P < 0.001). Incomplete scaffold apposition area was significantly lower if postdilatation was performed (0.01 ± 0.04 mm vs. 0.17 ± 0.11 mm , P < 0.001). Strut fractures could neither be observed before nor after postdilatation.

Conclusion: NC balloon postdilatation of Magmaris® BRS is required to achieve optimal expansion. It significantly reduces malapposition and can safely be performed without relevant rates of strut fracture.

Download full-text PDF

Source
http://dx.doi.org/10.1002/ccd.27794DOI Listing

Publication Analysis

Top Keywords

balloon postdilatation
16
magmaris® brs
12
postdilatation magnesium-based
8
brs implanted
8
diameter 321 ± 032 mm
8
postdilatation
7
brs
7
non-compliant balloon
4
magnesium-based bioresorbable
4
bioresorbable vascular
4

Similar Publications

Article Synopsis
  • Stent under-expansion can increase risks like neoatherosclerosis and stent thrombosis, especially in cases of severe coronary calcification, and intravascular lithotripsy (IVL) can help prevent this by ensuring proper stent expansion.
  • A 62-year-old male with unstable angina was treated for high-grade stenosis in the right coronary artery but experienced stent under-expansion despite multiple attempts to dilate it, including using ultra-high-pressure balloons.
  • The successful use of IVL led to complete stent expansion at a lower pressure, illustrating its effectiveness and safety in treating stent under-expansion, which is critical for preventing complications like restenosis and thrombosis.
View Article and Find Full Text PDF

Background: Few large-scale studies have evaluated the effectiveness of percutaneous coronary intervention (PCI) technological advances in the treatment of patients with unprotected left main coronary artery disease (LM-CAD). We aim to identify independent factors that affect the prognosis of PCI in patients with unprotected LM-CAD and to assess the impact of PCI technological advances on long-term clinical outcomes.

Methods And Results: A total of 4512 consecutive patients who underwent unprotected LM-CAD PCI at Fuwai Hospital from 2004 to 2016 were enrolled.

View Article and Find Full Text PDF

Introduction: Acute carotid-related stroke (CRS), with its large thrombo-embolic load and large volume of affected brain tissue, poses significant management challenges. First generation (single-layer) carotid stents fail to insulate the athero-thrombotic material; thus they are often non-optimized (increasing thrombosis risk), yet their use is associated with a significant (20-30%) risk of new cerebral embolism.

Aim: To evaluate, in a multi-center multi-specialty investigator-initiated study, outcomes of the MicroNET-covered (cell area ≈ 0.

View Article and Find Full Text PDF
Article Synopsis
  • Crohn's disease affects the small intestine, leading to complex strictures that are hard to diagnose and treat, prompting this study on how effective MRE and DBE-assisted EBD are in addressing these issues.!
  • The study looked at 10 patients who underwent DBE-assisted balloon dilatation after being diagnosed with strictures through MRE, showing that MRE accurately identified 75% of these strictures.!
  • With a high technical success rate of 72.2%, the study found that DBE was able to dilate most identified strictures without needing any surgical intervention during the follow-up period. !
View Article and Find Full Text PDF

Balloon venoplasty for right heart catheterization from the arm in patients with subclavian vein stenosis.

J Invasive Cardiol

September 2024

Department of Cardiovascular Disease, Memorial Healthcare System, Florida, USA.

Objectives: In the trans-radial era, arm venous access for right heart catheterization (RHC) is rising. Procedural success is affected by many factors, including subclavian/innominate vein stenosis (SVS) and pre-existing wires or catheters. In a study published previously by the same authors, 2% of cases had unsuccessful RHC through the arm, predominantly due to SVS.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!