Objectives: To evaluate hearing performance and safety measures in a large group of totally implantable active middle ear implant (AMEI) wearers.
Methods: Retrospective case review of 172 ears (166 patients) with sensorineural hearing loss undergoing implantation of a totally implantable AMEI. Pure-tone air and bone thresholds, pure-tone average, speech reception threshold (SRT), and word recognition scores (WRS) at phonetically balanced maximum and at 50 dB (WRS50) were assessed at baseline unaided, with a hearing aid (baseline aided [BLA]), with the implant at activation, and for best implant postoperative measurement. Adverse events were reviewed.
Results: Mean length of follow-up was 35.7 months. Compared to unaided, the implant provided significant gain at all frequencies through 6 kHz. Amount of gain was significantly better with the implant than BLA at 1,500 and 2,000 Hz (P ≤ .001). Implant SRT was also significantly improved compared to BLA (29.9 dB vs. 38.5 dB, P ≤ .001). At a 50-dB presentation level, WRS was significantly better with the implant than BLA (65.6% vs. 45.5%, P ≤ .001). Bone conduction thresholds were not affected by the implant. The most common adverse event was taste disturbance, which occurred in 39.1%. In subjects with preimplant tinnitus, 56.1% experienced resolution of their tinnitus. Revision surgery was required in 15.7%, and 4% were ultimately explanted.
Conclusion: The implant provided significant hearing benefit compared to unaided and to the subjects' own hearing aids in most test categories, with minimal adverse events. Many subjects experienced complete resolution of their tinnitus after implantation.
Level Of Evidence: 4 Laryngoscope, 128:2872-2878, 2018.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/lary.27246 | DOI Listing |
Alzheimers Dement
December 2024
Federal University of Rio Grande do Sul, Porto Alegre, Rio Grande do Sul, Brazil
Background: Alzheimer's disease (AD) is the leading cause of dementia worldwide and vascular dysfunction represents one of the first abnormalities in AD spectrum. Brain imaging techniques that use changes in hemodynamic signals to measure alterations in neurovascular coupling (NVC) have proven useful for early detection of cognitive deterioration. Pharmacological interventions targeting vascular risk factors, including simvastatin (SV), show promise in preventing dementia.
View Article and Find Full Text PDFCommun Med (Lond)
January 2025
MED-EL Elektromedizinische Geräte GmbH, Fürstenweg 77a, 6020, Innsbruck, Austria.
Background: Cochlear implants (CIs) are neuroprosthetic devices which restore hearing in severe-to-profound hearing loss through electrical stimulation of the auditory nerve. Current CIs use an externally worn audio processor. A long-term goal in the field has been to develop a device in which all components are contained within a single implant.
View Article and Find Full Text PDFBMC Oral Health
January 2025
Resident of Oral and Maxillofacial Radiology, Isfahan University of Medical Sciences, Isfahan, Iran.
Background: Early detection of peri-implant bone defects can improve long-term durability of dental implants. By the advances in cone-beam computed tomography (CBCT) scanners and introduction of new algorithms, it is important to find the most efficient protocol for detection of bone defects. This study aimed to assess the efficacy of metal artifact reduction (MAR) and advanced noise reduction (ANR) algorithms for detection of peri-implant bone defects.
View Article and Find Full Text PDFBiomol Biomed
January 2025
Department of medical oncology, Anhui Provincial People's Hospital, Hefei, China.
Totally implantable subcutaneous venous access ports (TISVAPs) are essential for long-term central venous chemotherapy, delivering medication directly into the central veins of patients. While they play a critical role in reducing patient discomfort, TISVAPs pose a notable risk of post-insertion infections-particularly concerning for oncology patients with compromised immune systems due to aggressive treatment regimens. Our research addresses this issue by developing a predictive nomogram to estimate the risk of TISVAP-associated infections.
View Article and Find Full Text PDFSensors (Basel)
December 2024
Department of Mechanical Engineering, University of Michigan, 2350 Hayward St., Ann Arbor, MI 48109, USA.
In this study, a method for determining the optimal location and orientation of an implantable piezoelectric accelerometer on the short process of the incus is presented. The accelerometer is intended to be used as a replacement for an external microphone to enable totally implantable auditory prostheses. The optimal orientation of the sensor and the best attachment point are determined based on two criteria-maximum pressure sensitivity sum and minimum loudness level sum.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!