Accuracy of remote continuous respiratory rate monitoring technologies intended for low care clinical settings: a prospective observational study.

Can J Anaesth

Department of Anesthesiology, Intensive Care and Emergency Medicine, University Medical Center Utrecht, Utrecht University, P.O. Box 85500, 3508 GA, Utrecht, The Netherlands.

Published: December 2018

Purpose: Altered respiratory rate (RR) has been identified as an important predictor of serious adverse events during hospitalization. Introduction of a well-tolerated continuous RR monitor could potentially reduce serious adverse events such as opioid-induced respiratory depression. The purpose of this study was to investigate the ability of different monitor devices to detect RR in low care clinical settings.

Methods: This was a prospective method-comparison study with a cross-sectional design. Thoracic impedance pneumography (IPG), frequency modulated continuous wave radar, and an acoustic breath sounds monitor were compared with the gold standard of capnography for their ability to detect RR in breaths per minute (breaths·min) in awake postoperative patients in the postanesthesia care unit. The Bland and Altman method for repeated measurements and mixed effect modelling was used to obtain bias and limits of agreement (LoA). Furthermore, the ability of the three devices to assist with correct treatment decisions was evaluated in Clarke Error Grids.

Results: Twenty patients were monitored for 1,203 min, with a median [interquartile range] of 61 [60-63] min per patient. The bias (98.9% LoA) were 0.1 (-7.9 to 7.9) breaths·min for the acoustic monitor, -1.6 (-10.8 to 7.6) for the radar, and -1.9 (-13.1 to 9.2) for the IPG. The extent to which the monitors guided adequate or led to inadequate treatment decisions (determined by Clarke Error Grid analysis) differed significantly between the monitors (P = 0.011). Decisions were correct 96% of the time for acoustic, 95% of the time for radar, and 94% of the time for IPG monitoring devices.

Conclusions: None of the studied devices (acoustic, IPG, and radar monitor) had LoA that were within our predefined (based on clinical judgement) limits of ± 2 breaths·min. The acoustic breath sound monitor predicted the correct treatment more often than the IPG and the radar device.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6244627PMC
http://dx.doi.org/10.1007/s12630-018-1214-zDOI Listing

Publication Analysis

Top Keywords

respiratory rate
8
low care
8
care clinical
8
serious adverse
8
adverse events
8
acoustic breath
8
correct treatment
8
treatment decisions
8
clarke error
8
breaths·min acoustic
8

Similar Publications

Utilizing Targeted Next-Generation Sequencing for Rapid, Accurate, and Cost-Effective Pathogen Detection in Lower Respiratory Tract Infections.

Infect Drug Resist

January 2025

Department of Respiratory and Critical Care Medicine, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, 430022, People's Republic of China.

Objective: To evaluate the diagnostic performance and clinical impact of targeted next-generation sequencing (tNGS) in patients with suspected lower respiratory tract infections.

Methods: Following propensity score matching, we compared the diagnostic performances of tNGS and metagenomic next-generation sequencing (mNGS). Furthermore, the diagnostic performance of tNGS was compared with that of culture, and its clinical impact was assessed.

View Article and Find Full Text PDF

Objective: This study aimed to investigate the levels of coagulation parameters in elderly patients with severe pneumonia and analyse their correlation with disease severity and prognosis.

Methods: A retrospective study was conducted on 207 elderly patients (aged ≥60 years) with severe pneumonia admitted to our hospital between January 2022 and December 2023. Demographic data, clinical characteristics and coagulation parameters, including prothrombin time (PT), activated partial thromboplastin time (APTT), thrombin time and fibrinogen (FIB), were collected.

View Article and Find Full Text PDF

Specific and sensitive detection of bovine coronavirus using CRISPR-Cas13a combined with RT-RAA technology.

Front Vet Sci

January 2025

Key Laboratory of Animal Medicine at Southwest Minzu University of Sichuan Province, College of Animal Science and Veterinary Medicine, Southwest Minzu University, Chengdu, China.

Introduction: Bovine coronavirus (BCoV) is an important pathogen of enteric and respiratory disease in cattle, resulting in huge economic losses to the beef and dairy industries worldwide. A specific and sensitive detection assay for BCoV is critical to the early-stage disease prevention and control.

Methods: We established a specific, sensitive, and stable assay for BCoV nucleic acid detection based on CRISPR/Cas13a combined with reverse transcription recombinase-aided amplification (RT-RAA) technology.

View Article and Find Full Text PDF

Purpose: This randomized, controlled trial aimed to assess the sedative effects of esketamine and sufentanil combined with propofol during EUS.

Patients And Methods: Three hundred and forty patients undergone EUS were randomly divided into two groups to receive esketamine 0.25 mg/kg combined with propofol (esketamine group) or sufentanil 0.

View Article and Find Full Text PDF

Background: Pulmonary arterial hypertension (PAH) poses a significant health challenge globally, with China experiencing a notable increase in its burden. Understanding the trends and factors contributing to PAH is crucial for developing effective public health strategies.

Methods: This study utilized data from the Global Burden of Disease (GBD) 2021 database to estimate the burden of PAH in China and worldwide from 1990 to 2021.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!