Background: Measuring bone mineral density (BMD) around acetabular prosthetic components with computed tomography (CT) is challenged by the complex anatomy and metal artifacts. Three-dimensional (3D) segmentation is required for the analysis, but it is usually not practically applicable on current CT workstations.

Purpose: To test the between-scan agreement and reliability of custom segmentation software for BMD measurements adjacent to cemented and uncemented acetabular cups in dual-energy CT (DECT).

Material And Methods: Twenty-four male patients with total hip arthroplasty were scanned and rescanned using 130-keV virtual monochromatic DECT images. Hemispherical regions of interest were defined slice-by-slice and BMD was calculated around the acetabular cup using custom segmentation software.

Results: In the uncemented cup, the mean BMD was 153 mg/cm with a between-scan difference of 10 mg/cm ( < 0.0001). In the cemented cup, the mean BMD was 186 mg/cm with a between-scan difference of 6 mg/cm ( = 0.15). In both uncemented and cemented cups the intraclass correlation coefficient between repeated measurements was >0.95 and narrow Bland-Altman Limits of Agreement.

Conclusion: BMD can be measured with high absolute between-scan agreement and good reliability adjacent to acetabular cemented and uncemented cups using DECT and segmentation software.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6117866PMC
http://dx.doi.org/10.1177/2058460118796539DOI Listing

Publication Analysis

Top Keywords

measurements adjacent
8
adjacent acetabular
8
acetabular cups
8
total hip
8
hip arthroplasty
8
between-scan agreement
8
custom segmentation
8
segmentation software
8
cemented uncemented
8
acetabular
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!