Background: In urology, lasers are used in a variety of endoscopic procedures such as ureteroscopy and retrograde renal surgery for stone fragmentation of urinary calculi and ablation of urothelial tumors. To perform these procedures, guidewires are used as a preliminary safe-mainstay for referencing the urinary tract. This study aims to determine the effect of two different lasers: holmium:YAG (Ho:YAG) and thulium:YAG (Tm:YAG) lasers on metal guidewires with PTFE coating (PTFE), nitinol guidewires with hydrophilic coating (Hydrophilic) and nitinol guidewires with hydrophilic listed coating (Zebra).

Methods: Different combinations of frequency (5, 10 and 12 Hz) and energy per pulse (0.5, 1.5, and 2.6 J) of Ho:YAG laser were applied on the three kinds of guidewires in two experiments (50 J vs. 100 J of total energy). For the Tm:YAG laser three power levels (5, 35, and 70 W) with a total energy of 100 J were applied to the guidewires. The degree of damage (0 to 5) of the guidewire was assessed after each laser application.

Results: A higher degree of injury of guidewires was related to higher values of total energy used for the Ho:YAG laser (P=0.036), and to higher values of power applied with the Tm:YAG (P=0.051). The most resistant guidewire to Ho:YAG laser energy was Zebra, followed by PTFE and Hydrophilic (P<0.001). With the Tm:YAG laser, PTFE guidewire appears to be the most resistant and the Hydrophilic the most fragile, although without reaching the statistical significance (P=0.223).

Conclusions: Both lasers revealed a harmful effect on the three tested guidewires. There was an association between the degree of injury and the amount of Ho:YAG laser energy and Tm:YAG laser power. The guidewire Zebra proved to be the safest when using Ho:YAG laser and the PTFE guidewire the most resistant to laser Tm:YAG. Further studies are necessary to confirm these results.

Download full-text PDF

Source
http://dx.doi.org/10.23736/S0393-2249.18.03050-3DOI Listing

Publication Analysis

Top Keywords

hoyag laser
12
total energy
12
nitinol guidewires
8
guidewires hydrophilic
8
higher values
8
guidewires
7
laser
6
energy
5
resistance guidewires
4
guidewires laser
4

Similar Publications

Ureteroscopic nephrolithotripsy is now being preferred over percutaneous nephrolithotomy for removal of kidney stones in children. Here, in this report, we have discussed persistent oxygen desaturation immediately after extubation in a two-year-old child who underwent Ho-YAG (holmium-yttrium-aluminium garnet) laser ureteroscopic nephrolithotripsy. The child developed bilateral pleural effusion after nephrolithotripsy and required continuous oxygen supplementation to maintain oxygen saturation above 95%, followed by ultrasound-guided thoracentesis.

View Article and Find Full Text PDF

Objectives: This study's aim is to evaluate the long-term quality of life and functional outcomes following cystoscopic excision of stress urinary incontinence (SUI) and pelvic organ prolapse (POP) mesh extruded into the urinary tract in women.

Patients And Methods: A retrospective chart review was performed of all cases of cystoscopic removal of extruded mesh at our high-volume tertiary care centre between April 2013 and August 2021. Postoperative patient-reported outcomes were collected via questionnaires: Urogenital Distress Inventory Short Form (UDI-6), EQ-5D-5L Visual analogue scale, ICIQ-Satisfaction (ICIQ-S) and additional questions regarding postoperative sexual function.

View Article and Find Full Text PDF

Ablation rates with Holmium:YAG and Thulium Fiber Laser: Influence of the stone phantom homogeneity. An in vitro study.

Prog Urol

August 2023

GRC n(o) 20, Groupe de recherche clinique sur la lithiase urinaire, hôpital Tenon, Sorbonne université, 75020 Paris, France; Service d'urologie, hôpital Tenon, Sorbonne université, Assistance publique-Hôpitaux de Paris, 4, rue de la Chine, 75020 Paris, France; PIMM, UMR 8006 CNRS-Arts et Métiers ParisTech, 151, boulevard de l'Hôpital, 75013 Paris, France. Electronic address:

Objectives: The lithotripsy efficiency (LE) in vitro study requires artificial or human stone samples (AS, HS). With the development of dusting lithotripsy, less ex vivo HS are available. We aimed to compare Thulium Fiber Laser (TFL) and Holmium:YAG (Ho:YAG)'s LE and define the most accurate LE parameter.

View Article and Find Full Text PDF

Objective: To describe the technique of transurethral harvesting of bladder mucosal graft using the Holmium:YAG (Ho-YAG) laser and describe the preliminary results from 7 cases where this graft was used for urethroplasty.

Materials And Methods: We performed a single-stage dorsal onlay urethroplasty using bladder mucosal graft in 7 patients with anterior urethral stricture. Transurethral harvesting was performed with the Ho-YAG laser.

View Article and Find Full Text PDF

Aim: The aim of this study was to analyze the outcomes of miniaturized nephrolithotomy (mini-perc) in the management of renal stones with a diameter smaller than 20 mm.

Materials And Methods: We retrospectively reviewed the records of 102 patients who underwent mini-perc between March 2015 and March 2020 in our department. The primary objective was the stone-free rate, but we also analyzed the retreatment rate, complications, hospital stay, operative time and reduction in hemoglobin level.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!