A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Reliability of 2D and 3D ultrasound for infant hip dysplasia in the hands of novice users. | LitMetric

Reliability of 2D and 3D ultrasound for infant hip dysplasia in the hands of novice users.

Eur Radiol

Department of Radiology and Diagnostic Imaging, 2A2.41 WC Mackenzie Health Sciences Centre, University of Alberta, 8440-112 Street, Edmonton, AB, T6G 2B7, Canada.

Published: March 2019

Purpose: Developmental dysplasia of the hip (DDH) diagnosis by two-dimensional ultrasound (2DUS) can have poor inter-rater reliability. 3D ultrasound (3DUS) may be more reliably performed, particularly by novice users. We compared intra- and inter-rater reliability between expert and novice operators performing 2DUS and 3DUS for DDH.

Materials And Methods: Infants with suspected DDH were assessed with 2DUS and 3DUS. Novice operators had 1.5 h of training and Experts had 5-15 years' experience. Images included two 2DUS static and two 3DUS sweep images per operator. Image quality was assessed by 5-point system (yes/no: full femoral head; full acetabular roof; horizontal iliac wing; os ischium; absent motion/artifact). 2DUS indices (alpha angle, coverage) were measured centrally by a blinded reader with 2 years DDH US experience. 3DUS was post-processed by semi-automated custom software generating acetabular surface models, indices and estimated probability of DDH. Gold-standard diagnosis of each hip as normal, borderline or dysplastic was based on radiologist review of expert 2DUS.

Results: Thirty infants, mean age 10.8 weeks were enrolled. Quality scores were 2.7±1.2 Novice versus 4.9±0.3 Expert for 2DUS (p = 0.04), and 4.2±1.0 Novice versus 4.9±0.3 Expert for 3DUS (p = 0.99). Inter-rater reliability was poor for 2DUS (ICC=0.10 for alpha angle, 0.04 for acetabular coverage) and moderate to high for 3DUS (ICC=0.73-0.83 for alpha angle, 0.55 for acetabular coverage). Intra-rater reliability and diagnostic accuracy was higher for 3DUS than 2DUS.

Conclusion: Novice operators can perform 3DUS for DDH with reliability and accuracy approaching expert sonographers. Novices perform 2DUS with poor reliability and accuracy.

Key Points: • Novice/expert inter-rater reliability improved from poor with 2DUS to moderate/high with 3DUS. • Novice operators using 3DUS correctly classified 57/58 (98%) of infant hips. • DDH can be reliably assessed by novice operators using 3DUS.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00330-018-5699-1DOI Listing

Publication Analysis

Top Keywords

novice operators
20
inter-rater reliability
16
3dus
12
alpha angle
12
novice
9
2dus
9
reliability
8
reliability ultrasound
8
novice users
8
2dus poor
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!