Influence of calibration on digital templating of hip arthroplasty.

Int Orthop

Department of Spine Surgery, Schön Klinik Düsseldorf, Düsseldorf, Germany.

Published: August 2019

Introduction: Digital templating for total joint replacement is the current standard. For image calibration, external calibration markers (ECM) are used. However, there are concerns regarding the precision of the method. This study aimed to identify the direct influence of calibration errors on digital templating.

Patients And Methods: A retrospective analysis of 100 post-operative radiographs with unilateral total hip arthroplasty was performed. The magnification factor of the ECM and of the internal prosthetic femoral head (ICM) as a reference value was calculated for each radiograph. Two blinded observers performed templating of the contralateral hip using a randomized list for all radiographs and both markers. The component size templated by the ECM magnification was compared to the reference by the ICM magnification.

Results: Mean magnification factors of ICM and ECM differed significantly (p = 0.006). The absolute difference was 5.2% (range 0.0-23.3%, SD 4.8%). Templating of the acetabular or the femoral component showed no significant differences (p = 0.120, p = 0.599). Differences of more than one size were found in 26% of the acetabular components and 14% of the femoral components and differences over two sizes in 10% respectively 3%. Correlation coefficients for magnification error and size differences of acetabular components were - 0.645 (p < 0.001) and for the femoral component - 0.607 (p < 0.001).

Interpretation: The calibration error of external calibration markers in digital templating for hip replacement influences component sizes significantly. Thus, correct positioning of ECM is of utmost importance.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00264-018-4120-7DOI Listing

Publication Analysis

Top Keywords

influence calibration
8
digital templating
8
hip arthroplasty
8
acetabular components
8
calibration digital
4
templating
4
templating hip
4
arthroplasty introduction
4
introduction digital
4
templating total
4

Similar Publications

Background The synthetic opioid tramadol is widely used as a pain reliever. Unlike other opioids, it is used freely worldwide, unaffected by international controls resulting in abuse and accidental intoxication. Analytical methods are necessary to prove tramadol abuse because 30% of the drug is excreted unchanged.

View Article and Find Full Text PDF

Machine Learning for Predicting Primary Graft Dysfunction After Lung Transplantation: An Interpretable Model Study.

Transplantation

January 2025

Department of Thoracic Surgery, The Affiliated Wuxi People's Hospital of Nanjing Medical University, Wuxi People's Hospital, Wuxi Medical Center, Nanjing Medical University, Wuxi, Jiangsu, China.

Background: Primary graft dysfunction (PGD) develops within 72 h after lung transplantation (Lung Tx) and greatly influences patients' prognosis. This study aimed to establish an accurate machine learning (ML) model for predicting grade 3 PGD (PGD3) after Lung Tx.

Methods: This retrospective study incorporated 802 patients receiving Lung Tx between July 2018 and October 2023 (640 in the derivation cohort and 162 in the external validation cohort), and 640 patients were randomly assigned to training and internal validation cohorts in a 7:3 ratio.

View Article and Find Full Text PDF

Derivation and validation of a prediction model for inadequate bowel preparation in Chinese outpatients.

Sci Rep

January 2025

Nursing Department, The First Affiliated Hospital, Zhejiang University School of Medicine, Building 17, 3rd Floor 79 Qingchun Road, Hangzhou, 310003, China.

The quality of bowel preparation is an important factor in the success of colonoscopy. However, multiple influencing factors that function together can lead to inadequate bowel preparation. The main objective of this study was to explore the specific factors that affect the quality of bowel preparation, with the goal of deriving and validating a predictive model for inadequate bowel preparation in Chinese outpatients.

View Article and Find Full Text PDF

Introduction of curvature index and improvement of calibration curves used in ion chromatography (IC), atomic absorption spectrometry (AAS), UV-vis spectrophotometry and catalytic combustion.

Anal Chim Acta

January 2025

Wastewater Laboratory Targoviste, Water Company Targoviste-Dambovita, Calea Ploiesti Street, 130 145, Targoviste, Romania. Electronic address:

Background: The calibration curve represents the relationship between known CRM and the response of the instrument (e.g absorbance). Most of the time, in analytical chemistry standards, the linear function (of the 1st degree) is used to characterize the calibration curves.

View Article and Find Full Text PDF

Objectives: Sepsis, a critical condition caused by a dysregulated host response to infection, has high morbidity and mortality rates. Timely diagnosis and treatment are vital for improving patient outcomes. This study explores the potential role of CXCL5 in the diagnosis, severity assessment, and prognosis of sepsis.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!