Background And Aims: Laryngeal mask airway (LMA) CTrach™ and Airtraq videolaryngoscopes are useful for endotracheal intubation in patients with limited cervical spine movements and other predicted difficult airways. We aimed to compare LMA CTrach™ and Airtraq videolaryngoscopes as conduits for endotracheal intubation in patients with simulated limitation of cervical spine movements by manual in-line stabilization (MILS).

Material And Methods: This was a prospective, randomized study including 50 patients undergoing elective surgeries under general anesthesia. Patients were assigned to undergo intubation using Airtraq ( = 25) or LMA CTrach™ ( = 25) by an experienced anesthesiologist, while MILS was provided. Laryngoscopy and intubation were compared in terms of time taken to obtain optimal laryngeal view, successful intubation, total time, percentage of glottis opening (POGO) score, maneuvers required for optimal laryngeal view and alignment of endotracheal tube, and number of attempts and complications. An integrated score was calculated to classify the attempt as good, restricted, or poor.

Results: Time taken to obtain optimal laryngeal view, successful intubation, and total time in both groups were comparable. POGO score >50% was seen in 25 and 21 patients in Groups A and C. Seventy-six percent and ninety-six percent in Groups A and C, respectively, had a good integrated score; 6% and 1% had restricted score; none had a poor score; and the difference between them was statistically significant ( = 0.042).

Conclusions: LMA CTrach™ and Airtraq are similar with respect to time taken for obtaining optimal laryngeal view, successful intubation, and total time when used for intubation in patients with simulated limitation of cervical spine movements.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6066898PMC
http://dx.doi.org/10.4103/joacp.JOACP_330_16DOI Listing

Publication Analysis

Top Keywords

ctrach™ airtraq
16
intubation patients
16
cervical spine
16
spine movements
16
lma ctrach™
16
optimal laryngeal
16
laryngeal view
16
airtraq videolaryngoscopes
12
endotracheal intubation
12
patients simulated
12

Similar Publications

Background: Videolaryngoscopes increased the safety of patients requiring endotracheal intubation. Their particular beneficiaries are healthcare providers inexperienced in intubation. We compared the learning curve of McGrath, UESCOPE, and Airtraq VLs among the first-year anesthesiology residents.

View Article and Find Full Text PDF
Article Synopsis
  • There is a lack of consensus on the best videolaryngoscope for awake tracheal intubation in patients expected to have difficult airways, leading to a study comparing the C-MAC D-Blade® and Airtraq®.
  • A randomized clinical trial involved 90 adult patients undergoing surgery with anticipated difficult airways, focusing on the first-attempt tracheal intubation success rate as the primary outcome.
  • Results showed a significantly higher first-attempt success rate for the C-MAC D-Blade (84%) compared to the Airtraq (62%), while overall success rates were similar for both, suggesting the C-MAC D-Blade may be more effective for initial intubation attempts.
View Article and Find Full Text PDF

Background And Aim: Intubation with Macintosh requires flexing the lower cervical spine and extending the atlanto-occipital joint to create a "line of sight." Primary aim of study was to compare the extent of cervical spine movement during laryngoscopy using conventional Macintosh laryngoscope and Airtraq.

Material And Methods: A total of 25 patients of either sex between the age group of 18 and 60 years, having American Society of Anesthesiologists (ASA) physical status of Grade-I and Grade-II, scheduled for elective surgery under image control requiring general anesthesia and intubation were enrolled.

View Article and Find Full Text PDF

Severe traumatic damage to the brain-to-body signaling bundle that results in bruising and a partial or total tear of the spinal cord is known as a spinal cord injury (SCI). SCI may develop at the time of an event or after. It can also develop while handling the patient and can worsen during the transportation of the patient.

View Article and Find Full Text PDF

The aim of the study was to test the hypothesis that the results obtained with three different types of video laryngoscopes (UESCOPE VL-400, I-View, Non-Channeled Aitraq) with and without an endotracheal stylet should be better than the results obtained with a Macintosh laryngoscope in a simulated out-of-hospital scenario by a person without clinical experience. Primary outcome measures were the time taken to successfully achieve tracheal intubation (TI). Secondary outcomes included the grade of glottic view (Cormack and Lehane grades 1-4), the incidence of successful TI, the number of audible dental clicks indicating potential dental damage, the level of effort required to perform TI, and the operator's comfort during the procedure.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!