Aims: The purpose of this work is to present and evaluate the performance of a novel Automatic Lab-in-vial Escherichia coli Remote Tracking technology based on an automated real-time defined substrate approach, implemented in both portable and in situ instruments.

Methods And Results: We present the fresh water calibration procedure, and assess performance using side-by-side comparison with most probable number (MPN) approaches in terms of accuracy, reproducibility and capability to correctly generate early-warning alerts. Long-term data from an operational in situ deployment at a potential bathing site is presented as well.

Conclusions: Automatic Lab-in-vial Escherichia coli Remote Tracking technology is shown to be an accurate and rapid bacterial quantification technology, capable of autonomous in situ measurements with metrological capabilities comparable to those of an approved laboratory using MPN microplate techniques.

Significance And Impact Of The Study: Rapid quantification of bacterial pollution is a requirement in water quality applications ranging from recreational water use, agriculture and aquaculture to drinking and wastewater treatment. The method and instruments presented in this work should enable fast and accurate bacterial concentration measurements to be performed in a portable or in situ manner, thus simplifying operational logistics, reducing time-to-result delays, and eliminating sample transportation constraints associated with traditional techniques.

Download full-text PDF

Source
http://dx.doi.org/10.1111/jam.14066DOI Listing

Publication Analysis

Top Keywords

escherichia coli
12
automatic lab-in-vial
8
lab-in-vial escherichia
8
coli remote
8
remote tracking
8
tracking technology
8
portable in situ
8
autonomous system
4
system rapid
4
rapid field
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!