Background: A virtual guidance framework is used to assist the conventional method of virtual implant planning system (VIPS). The study null hypothesis was that its screw placement accuracy is equal to that of conventional VIPS.
Methods: In 34 distal radius sawbone models, 3D implant planning was performed. A camera attached to the surgical drill was used to support screw positioning. Differences of angles/tip distances between planned and placed screws were identified in intraoperative cone beam Computer tomography (CT) and compared with already existing data from 22 patients treated by conventional VIPS.
Results: The virtual guidance group showed tip distances of 1.02 ± 0.56 mm, azimuth of 3.69° ± 4.34°, and inclination of 1.75° ± 1.37°, whereas the VIPS group showed tip distances of 2.23 ± 0.99 mm (P < 0.001), azimuth of 23.17° ± 33.50° (P < 0.001), and inclination angle of 4.18° ± 6.29° (P = 0.001).
Conclusions: The results reveal that using a guidance framework leads to a higher accuracy in screw placement compared with the conventional VIPS itself.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/rcs.1945 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!