Aims: The aim of this study was to compare the strut coverage of the XIENCE stent with that of the BuMA Supreme sirolimus-eluting cobalt-chromium stent, which has a shorter drug elution, on optical coherence tomography (OCT) one or two months after implantation.

Methods And Results: The PIONEER-II OCT trial was a multicentre, two-arm randomised trial, which comprised two cohorts: cohort-1 underwent an OCT imaging one month after coronary intervention (BuMA: 16 patients with 18 lesions, XIENCE: 15 patients with 17 lesions), whereas cohort-2 underwent OCT at two months (BuMA: 21 patients with 21 lesions, XIENCE: 23 patients with 28 lesions). The primary hypotheses were non-inferiority of the BuMA stent to the XIENCE stent in percent strut coverage at one month (cohort-1) or two months (cohort-2). In cohort-1, the BuMA stent was non-inferior to the XIENCE stent in terms of the strut coverage (83.8±10.4% for BuMA vs. 73.0±17.5% for XIENCE, pfor noninferiority <0.001), and was also significantly higher than the XIENCE (pfor superiority 0.037). In cohort-2, the BuMA stent was non-inferior to the XIENCE stent in OCT strut coverage (80.3±18.3% vs. 73.3±21.3%, pfor noninferiority 0.006, pfor superiority 0.24). Healing scores showed better healing in the BuMA stent in cohort-1 (32.36±21.59 vs. 54.88±34.65, p=0.027), whereas there was comparable healing between the BuMA and XIENCE stents in cohort-2 (39.86±37.77 vs. 53.75±42.84, p=0.25).

Conclusions: The BuMA Supreme had a faster coverage than the XIENCE at one month, presumably due to faster and shorter sirolimus elution. The difference in tissue coverage became less evident at two months.

Download full-text PDF

Source
http://dx.doi.org/10.4244/EIJ-D-18-00461DOI Listing

Publication Analysis

Top Keywords

xience stent
16
patients lesions
16
strut coverage
12
buma supreme
8
stent
8
stent xience
8
pioneer-ii oct
8
oct months
8
underwent oct
8
buma patients
8

Similar Publications

Background: Long-term clinical data on biodegradable-polymer (BP) drug-eluting stents (DES) are limited. The objective of this study was to assess the long-term safety and efficacy of the BP-DES SYNERGY compared to XIENCE V, a durable-polymer (DP)-DES.

Methods: We compared patients treated with BP-DES or DP-DES at our center from 2008 to 2020.

View Article and Find Full Text PDF

Objectives: Percutaneous old balloon angioplasty is still the preferred treatment for the treatment of below-the-knee (BTK) arteries in chronic limb-threatening ischemia (CLTI). In the case of a suboptimal angioplasty result, a bailout stenting is required. So far, few data are available to assess the outcomes of bailout stenting after BTK angioplasty.

View Article and Find Full Text PDF
Article Synopsis
  • - The NEXT trial compared the Nobori biolimus-eluting stent (BES) and the XIENCE/PROMUS everolimus-eluting stent (EES) by analyzing data from 3,235 patients over 8-12 months at 18 centers.
  • - Researchers examined various cross-sections before and after stent implantation, measuring volume changes in different tissue types at plaque sites, finding no significant differences in vascular response between the two stent types after 8 months.
  • - Despite the lack of significant findings and the trial being underpowered for comprehensive statistical analysis, the results indicated similar vascular tissue reactions for both stents, despite differences in material and thickness.
View Article and Find Full Text PDF
Article Synopsis
  • - This study developed a machine learning model to personalize the duration of dual antiplatelet therapy (DAPT) after coronary stent implantation, aiming to reduce bleeding risks associated with prolonged therapy.
  • - The model, tested on data from several clinical trials, indicated that 84.5% of participants could safely receive a shorter 3-month DAPT without significantly increasing the risk of major adverse cardiovascular events.
  • - Validation results showed that patients on abbreviated DAPT had a lower incidence of major bleeding compared to those on the standard 12-month DAPT, highlighting the effectiveness of personalized therapy in improving patient outcomes.
View Article and Find Full Text PDF

Background: Angina pectoris can occur in up to 40% of patients following percutaneous coronary intervention (PCI). There is limited data assessing whether the type of stent implanted during revascularization can predict post-PCI angina symptoms.

Methods: In this study, data regarding revascularization characteristics including the stent type in patients admitted for PCI was collected.

View Article and Find Full Text PDF