Acceptability, Accuracy, and Safety of Disposable Transnasal Capsule Endoscopy for Barrett's Esophagus Screening.

Clin Gastroenterol Hepatol

National Institute for Health Research Biomedical Research Center in Gastrointestinal and Liver Diseases at Nottingham University Hospitals NHS Trust and the University of Nottingham, Queens Medical Center Campus, Nottingham, United Kingdom. Electronic address:

Published: March 2019

Background & Aims: Screening for Barrett's esophagus (BE) with conventional esophagogastroduodenoscopy (C-EGD) is expensive. We assessed the performance of a clinic-based, single use transnasal capsule endoscope (EG Scan II) for the detection of BE, compared to C-EGD as the reference standard.

Methods: We performed a prospective multicenter cohort study of patients with and without BE recruited from 3 referral centers (1 in the United States and 2 in the United Kingdom). Of 200 consenting participants, 178 (89%) completed both procedures (11% failed EG Scan due to the inability to intubate the nasopharynx). The mean age of participants was 57.9 years and 67% were male. The prevalence of BE was 53%. All subjects underwent the 2 procedures on the same day, performed by blinded endoscopists. Patients completed preference and validated tolerability (10-point visual analogue scale [VAS]) questionnaires within 14 days of the procedures.

Results: A higher proportion of patients preferred the EG Scan (54.2%) vs the C-EGD (16.7%) (P < .001) and the EG Scan had a higher VAS score (7.2) vs the C-EGD (6.4) (P = .0004). No serious adverse events occurred. The EG Scan identified any length BE with a sensitivity value of 0.90 (95% CI, 0.83-0.96) and a specificity value of 0.91 (95% CI, 0.82-0.96). The EG Scan identified long segment BE with a sensitivity value of 0.95 and short segment BE with a sensitivity values of 0.87.

Conclusions: In a prospective study, we found the EG Scan to be safe and to detect BE with higher than 90% sensitivity and specificity. A higher proportion of patients preferred the EG Scan to C-EGD. This device might be used as a clinic-based tool to screen populations at risk for BE. ISRCTN registry identifier: 70595405; ClinicalTrials.gov no: NCT02066233.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6330075PMC
http://dx.doi.org/10.1016/j.cgh.2018.07.019DOI Listing

Publication Analysis

Top Keywords

transnasal capsule
8
barrett's esophagus
8
scan
8
higher proportion
8
proportion patients
8
patients preferred
8
preferred scan
8
scan identified
8
segment sensitivity
8
c-egd
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!