Reliability of Hand-Held Transcranial Doppler with M-mode Ultrasound in Middle Cerebral Artery Measurement.

J Med Ultrasound

Unidad de Neurología Vascular, Servicio de Neurología, Departamento de Neurología y Psiquiatría, Clínica Alemana de Santiago, Facultad de Medicina Clínica Alemana Universidad del Desarrollo, Santiago, Chile.

Published: January 2017

Purpose: To determine the intra- and interrater agreement of mean flow velocity (MFV) and pulsatility index (PI) measurement in middle cerebral arteries, assessed by transcranial Doppler (TCD) with M-mode.

Methods: Masked experienced neurosonologists performed TCD with M-mode using handheld probe in healthy adult volunteers. The Bland-Altman method for concordance and intraclass correlation coefficient were used.

Results: Seventy-seven healthy volunteers and seven raters participated (3 on regular TCD shift and 4 off-shift). The intrarater absolute mean difference between measurements was 5.5 cm/s [95% confidence interval (CI), 4.7-6.3] for MVF and 0.073 (95% CI, 0.063-0.083) for PI. The difference between MFV measurements was significantly higher in off-shift raters ( = 0.015). The interrater absolute mean difference between measurements was 6.5 cm/s (95% CI, 5.5-7.5) for MVF and 0.065 (95% CI, 0.059-0.071) for PI. No influence was found for the middle cerebral artery side, volunteer's sex, or age, and there was no significant difference between raters. The intraclass correlation coefficient was 82.2% (95% CI 77.8-85.6) and 72.9% (95% CI 67.4-77.6) for MFV and PI, respectively.

Conclusions: There exists good intra- and interrater agreement in MFV and PI measurements using M-mode TCD. These results support the use of this noninvasive tool and are important for clinical and investigational purposes.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6029317PMC
http://dx.doi.org/10.1016/j.jmu.2016.12.001DOI Listing

Publication Analysis

Top Keywords

middle cerebral
12
transcranial doppler
8
cerebral artery
8
intra- interrater
8
interrater agreement
8
intraclass correlation
8
correlation coefficient
8
absolute difference
8
difference measurements
8
measurements cm/s
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!