Studies towards hcTnI Immunodetection Using Electrochemical Approaches Based on Magnetic Microbeads.

Sensors (Basel)

Nanobiotechnology for Diagnostics (Nb4D), Department of Chemical and Biomolecular Nanotechnology, Institute for Advanced Chemistry of Catalonia (IQAC) of the Spanish Council for Scientific Research (CSIC), Jordi Girona 18-26, 08034 Barcelona, Spain.

Published: July 2018

Different electrochemical strategies based on the use of magnetic beads are described in this work for the detection of human cardiac troponin I (hcTnI). hcTnI is also known as the gold standard for acute myocardial infarction (AMI) diagnosis according to the different guidelines from the European Society of Cardiology (ESC) and the American College of Cardiology (ACC). Amperometric and voltamperometric sandwich magnetoimmunoassays were developed by biofunctionalization of paramagnetic beads with specific antibodies. These bioconjugates were combined with biotinylated antibodies as detection antibodies, with the aim of testing different electrochemical transduction principles. Streptavidin labeled with horseradish peroxidase was used for the amperometric magnetoimmunoassay, reaching a detectability of 0.005 ± 0.002 µg mL in 30 min. Cadmium quantum dots-streptavidin bioconjugates were used in the case of the voltamperometric immunosensor reaching a detectability of 0.023 ± 0.014 µg mL.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6111321PMC
http://dx.doi.org/10.3390/s18082457DOI Listing

Publication Analysis

Top Keywords

based magnetic
8
reaching detectability
8
studies hctni
4
hctni immunodetection
4
immunodetection electrochemical
4
electrochemical approaches
4
approaches based
4
magnetic microbeads
4
microbeads electrochemical
4
electrochemical strategies
4