Objectives: Optical coherence tomography (OCT) can noninvasively visualize in vivo tissue microstructure with high spatial resolution that approaches the histologic level. Currently, OCT studies in gynecology are few and limited to a conventional 1.3 μm center wavelength swept light source which provides high spatial resolution but limited penetration depth. Here, we present a novel endoscopic OCT system with improved penetration depth and high resolution.

Methods: A novel endoscopic OCT system was developed based on a 1.7 µm swept source laser, which is capable of deeper tissue penetration due to its longer wavelength. To evaluate the performance of system, we imaged the human vaginas in vivo with both conventional 1.3 and 1.7 μm endoscopic OCT systems.

Results: With the 1.7 μm endoscopic OCT system, imaging depth was improved by more than 25%, allowing better visualization of the lamina propria and clear contrast of the epithelial layer from the surrounding tissues.

Conclusion: The significantly improved performance of the novel 1.7 μm OCT imaging system demonstrates its potential use as a minimally-invasive monitoring tool of vaginal health in gynecologic practice. Lasers Surg. Med. 51:120-126, 2019. © 2018 Wiley Periodicals, Inc.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6353703PMC
http://dx.doi.org/10.1002/lsm.23003DOI Listing

Publication Analysis

Top Keywords

endoscopic oct
16
oct system
12
optical coherence
8
coherence tomography
8
high spatial
8
spatial resolution
8
penetration depth
8
novel endoscopic
8
17 μm endoscopic
8
oct
7

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!