Article Abstract

Purpose: To construct a predictive model that describes how the duration and symmetry of a k-space-weighted image contrast (KWIC) window affects the temporal resolution of differently sized ultrasound foci when using a pseudo-golden angle stack-of-stars acquisition.

Methods: We performed a modulation analysis of proton resonance frequency temperature measurements to create the temporal modulation transfer function for KWIC windows of different symmetry and temporal duration. We reconstructed simulated ultrasound heating trajectories and stack-of-stars k-space data as well as experimental phantom data using the same trajectories. Images were reconstructed using symmetric and asymmetric KWIC windows of 3 different temporal durations. Simulated results were compared against temporal modulation transfer function predictions, experimental results, and the original simulated temperatures.

Results: The temporal modulation transfer function shows that temporal resolution with KWIC reconstructions depend on the object size. The KWIC window duration affected SNR and severity of undersampling artifacts. Accuracy and response delay improved as the KWIC window duration decreased or the size of the heated region within the KWIC plane increased. Precision worsened as the window duration decreased. Using a symmetric window eliminated the response delay to heated region size but introduced a large reconstruction delay.

Conclusion: The accuracy and precision of proton resonance frequency temperature measurements from a stack-of-stars acquisition using a sliding KWIC window reconstruction are dependent on the size of the KWIC window and the size and shape of the heated region. The temporal modulation transfer function of KWIC reconstructions for any object size can predict the temporal response to changes in signal being acquired, such as temperature and contrast enhancement.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6258296PMC
http://dx.doi.org/10.1002/mrm.27383DOI Listing

Publication Analysis

Top Keywords

kwic window
20
temporal modulation
16
modulation transfer
16
transfer function
16
proton resonance
12
resonance frequency
12
window duration
12
heated region
12
kwic
10
temporal
9

Similar Publications

Article Synopsis
  • The study aims to create a model that links the duration and symmetry of k-space-weighted image contrast (KWIC) windows to the temporal resolution in ultrasound imaging.
  • A modulation analysis of temperature measurements was conducted, leading to the comparison of simulated and experimental results using different KWIC window setups.
  • Findings reveal that the effectiveness of KWIC reconstructions varies with object size, with shorter window durations improving accuracy but potentially reducing precision; symmetric windows minimize response delays but complicate reconstruction timings.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!