Purpose: The aim of this study was to evaluate the ability of computer-aided detection (CAD) and human readers to detect pulmonary nodules ≥5 mm using 100 kV ultra-low-dose computed tomography (ULDCT) utilizing a tin filter.

Materials And Methods: After informed consent, 55 patients prospectively underwent standard-dose chest CT (SDCT) using 120 kV followed by ULDCT using 100 kV/tin. Reference nodules ≥5 mm were identified by a thoracic radiologist using SDCT. Four thoracic radiologists marked detected nodules on SDCT and ULDCT examinations using a dedicated computer workstation. After a 6-month memory extinction, readers were shown the same ULDCT cases with all CAD markings as well as their original detections, and characterized CAD detections as true positive or false positive.

Results: Volume CT Dose index (CTDIvol) for SDCT and ULDCT were 5.3±2 and 0.4±0.2 mGy (P<0.0001), respectively. Forty-five reference nodules were detected in 30 patients. Reader sensitivity varied widely but similarly for SDCT (ranging from 45% to 87%) and ULDCT (45% to 83%). CAD sensitivity was 76% (34/45) for SDCT and 71% (32/45) for ULDCT. After CAD, reader sensitivity substantially improved by 19% and 18% for 2 readers, and remained nearly unchanged for the other 2 readers (0% and 2%), despite reader perception that many more nodules were identified with CAD. There was a mean of 2 false-positive CAD detections/case.

Conclusions: ULDCT with 100 kV/tin reduced patient dose by over 90% without compromising pulmonary nodule detection sensitivity. CAD can substantially improve nodule detection sensitivity at ULDCT for some readers, maintaining interobserver performance.

Download full-text PDF

Source
http://dx.doi.org/10.1097/RTI.0000000000000348DOI Listing

Publication Analysis

Top Keywords

nodules ≥5 mm
8
sdct uldct
8
uldct
5
prospective pilot
4
pilot evaluation
4
evaluation radiologists
4
radiologists computer-aided
4
computer-aided pulmonary
4
pulmonary nodule
4
nodule detection
4

Similar Publications

The prevalence of goiter, thyroid nodules, and thyroid cancers in the pediatric population has increased. In some rare cases, local conditions such as juvenile desmoid-type fibromatosis (JDTF) can mimic specific thyroid pathology, complicating the diagnostic process. A 17-year-old obese adolescent girl was admitted to the Endocrinology Department with progressive swelling on the left side of the neck, persisting for approximately one year, recently accompanied by dysphonia and inspiratory dyspnea, and ultimately diagnosed as a unilateral nodular goiter associated with compressive phenomena.

View Article and Find Full Text PDF

With technological development, ionizing radiation has found applications in numerous occupations. However, the determination and quantification of the damage resulting from exposure to it remains rather unclear, along with the damage to particular organs. The aim of this systematic review was to investigate the relationship between low-dose ionizing radiation (LDIR) in exposed workers and possible functional changes and cancer development in the thyroid gland.

View Article and Find Full Text PDF

The detection of unexpected findings (UF) during CT scans of patients undergoing TAVR is frequent; however, it is unclear whether such findings have a clinical impact on the TAVR pathway. We conducted a retrospective, single-center observational study enrolling patients who were candidates for TAVR. All enrolled patients underwent a CT scan before valve implantation.

View Article and Find Full Text PDF

Lung cancer is the primary cause of cancer-related deaths. Most patients are typically diagnosed at advanced stages. Low-dose computed tomography (LDCT) has been proven to reduce lung cancer mortality, but screening programs using LDCT are associated with a high number of false positives and unnecessary thoracotomies.

View Article and Find Full Text PDF

Background And Objectives: Lung cancer screening is critical for early detection and management, particularly through the use of computed tomography (CT). This study aims to compare the Lung Imaging Reporting and Data System (Lung-RADS) Version 2022 with the British Thoracic Society (BTS) guidelines in classifying solid pulmonary nodules detected at lung cancer screening CT examinations.

Materials And Methods: This retrospective study included 224 patients who underwent lung cancer screening CT between 2016 and 2022 and had a reported solid pulmonary nodule.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!