Introduction: Epidural anesthesia is frequently used to provide postoperative analgesia following major surgical procedures. Secure fixation of the epidural catheter is necessary to prevent premature dislodgment and loss of epidural analgesia. Using an in vitro model, the current prospective study evaluates different types of dressings for securement of an epidural catheter by quantifying the force in Newtons (N) required for dislodgment using a digital force gage.

Methods: Four methods of epidural catheter securement were used on a simulator mannequin: 1) Suresite Window Clear Dressing, 2) Op-Site Post-Op Visible Dressing, 3) Steri-Strips and Suresite Window Clear Dressing, and 4) Steri-Strips and Op-Site Post-Op Visible Dressing. Each method of securement was assessed 10 times to calculate the mean force required to dislodge the catheter. Mean force of dislodgment for each method was compared using parametric tests.

Results: The force (mean ± SD) required for catheter dislodgment for the four methods was 14.0±2.9, 2, 10.7±1.5, 8.6±2.3, and 9.6±2.2 N, respectively. The pairwise difference showed that the Suresite Window Clear Dressing was the best securement method when compared with other methods.

Conclusion: Our study demonstrates the advantage of the Suresite Window Clear Dressing in securing the epidural catheter. Future clinical trials are needed to validate these findings.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6054289PMC
http://dx.doi.org/10.2147/LRA.S172799DOI Listing

Publication Analysis

Top Keywords

epidural catheter
16
suresite window
16
window clear
16
clear dressing
16
securement epidural
8
op-site post-op
8
post-op visible
8
visible dressing
8
dressing steri-strips
8
force required
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!