Background: Assessment of percent body fat (%BF) is typically measured with bioelectrical impedance (BIA) as a proxy for dual-energy X-ray absorptiometry (DXA). Notably, poorer agreement between BIA and DXA among persons who are overweight or obese has been reported. The use of electrical impedance myography (EIM) as a proxy for DXA has not been validated. The objective was to evaluate an EIM device and two multi-frequency BIA devices with the reference standard (DXA) stratified by weight status and gender.

Methods: In a convenience sample of 82 adults, %BF assessed by EIM and two BIA devices was compared to DXA. Agreement between devices was tested with intra-class correlation coefficients (ICC) and Bland-Altman plots.

Results: Agreement between DXA and EIM (ICC=0.77) was poorer than the agreement between either BIA device with DXA (ICC>0.87). Stratified by sex, agreement between EIM and DXA was greater for men than women (ICC=0.81 and ICC=0.61, respectively). Stratified by BMI, agreement between EIM and DXA was best for normal-weight individuals (ICC=0.89) and progressively poorer for overweight (ICC=0.80) and obese (ICC=0.67) individuals. Bland-Altman plots revealed wide limits of agreement and an increase in EIM mean difference as average %BF increased. Similar trends were seen in BIA assessments.

Conclusions: EIM and BIA substantially underestimate %BF in overweight and obese individuals. Wide limits of agreement coupled with variable ICC limit device interchangeability with one another and limit clinical utility.

Download full-text PDF

Source
http://dx.doi.org/10.23736/S0022-4707.18.08505-5DOI Listing

Publication Analysis

Top Keywords

dxa
9
electrical impedance
8
impedance myography
8
percent body
8
body fat
8
dual-energy x-ray
8
x-ray absorptiometry
8
agreement
8
poorer agreement
8
agreement bia
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!