Purpose: To design a forward view antenna for prostate imaging at 7 T, which is placed between the legs of the subject in addition to a dipole array.

Materials And Methods: The forward view antenna is realized by placing a cross-dipole antenna at the end of a small rectangular waveguide. Quadrature drive of the cross-dipole can excite a circularly polarized wave propagating along the axial direction to and from the prostate region. Functioning of the forward view antenna is validated by comparing measurements and simulations. Antenna performance is evaluated by numerical simulations and measurements at 7 T.

Results: Simulations of B on a phantom are in good correspondence with measurements. Simulations on a human model indicate that the signal-to-noise ratio (SNR), specific absorption rate (SAR) efficiency and SAR increase when adding the forward view antenna to a previously published dipole array. The SNR increases by up to 18% when adding the forward view antenna as a receive antenna to an eight-channel dipole array in vivo.

Conclusions: A design for a forward view antenna is presented and evaluated. SNR improvements up to 18% are demonstrated when adding the forward view antenna to a dipole array.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6175442PMC
http://dx.doi.org/10.1002/nbm.3993DOI Listing

Publication Analysis

Top Keywords

forward view
32
view antenna
32
design forward
12
adding forward
12
dipole array
12
antenna
11
view
8
antenna prostate
8
prostate imaging
8
imaging 7 t
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!