Background: Intra-fraction motion represents a crucial issue in the era of precise radiotherapy in several settings, including breast irradiation. To date, only few data exist on real-time measured intra-fraction motion in breast cancer patients. Continuous surface imaging using visible light offers the capability to monitor patient movements in three-dimensional space without any additional radiation exposure. The aim of the present study was to quantify the uncertainties of possible intra-fractional motion during breast radiotherapy.
Material And Methods: One hundred and four consecutive patients that underwent postoperative radiotherapy following breast conserving surgery or mastectomy were prospectively evaluated during 2028 treatment sessions. During each treatment session the patients' motion was continuously measured using the Catalyst™ optical surface scanner (C-RAD AB, Sweden) and compared to a reference scan acquired at the beginning of each session. The Catalyst system works through an optical surface imaging with light emitting diode (LED) light and reprojection captured by a charge coupled device (CCD) camera, which provide target position control during treatment delivery with a motion detection accuracy of 0.5 mm. For 3D surface reconstruction, the system uses a non-rigid body algorithm to calculate the distance between the surface and the isocentre and using the principle of optical triangulation. Three-dimensional deviations and relative position differences during the whole treatment fraction were calculated by the system and analyzed statistically.
Results: Overall, the maximum magnitude of the deviation vector showed a mean change of 1.93 mm ± 1.14 mm (standard deviation [SD]) (95%-confidence interval: [0.48-4.65] mm) and a median change of 1.63 mm during dose application (beam-on time only). Along the lateral and longitudinal axis changes were quite similar (0.18 mm ± 1.06 mm vs. 0.17 mm ± 1.32 mm), on the vertical axis the mean change was 0.68 mm ± 1.53 mm. The mean treatment session time was 154 ± 53 (SD) seconds and the mean beam-on time only was 55 ± 16 s. According to Friedman's test differences in the distributions of the three possible directions (lateral, longitudinal and vertical) were significant (p < 0.01), in post-hoc analysis there were no similarities between any two of the three directions.
Conclusion: The optical surface imaging system is an accurate and easy tool for real-time motion management in breast cancer radiotherapy. Intra-fraction motion was reported within five millimeters in all directions. Thus, intra-fraction motion in our series of 2028 treatment sessions seems to be of minor clinical relevance in postoperative radiotherapy of breast cancer.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6048710 | PMC |
http://dx.doi.org/10.1186/s13014-018-1072-4 | DOI Listing |
Cureus
November 2024
Medical Physics, BC Cancer Kelowna, Kelowna, CAN.
Linac-based stereotactic radiosurgery (SRS) with planning target volume (PTV) margins <1 mm has become increasingly common in recent years. Optical surface imaging for surface-guided radiation therapy (SGRT) is often used for intra-fraction motion monitoring during these treatments to facilitate the use of a smaller PTV margin by providing real-time quantitative patient positioning information. However, rotating the couch introduces errors to SGRT-reported translations and rotations that can be problematic for SRS treatments with non-coplanar arcs and very small PTV margins.
View Article and Find Full Text PDFTech Innov Patient Support Radiat Oncol
December 2024
Applied Radiation Therapy Trinity (ARTT), Discipline of Radiation Therapy, School of Medicine, Trinity St. James's Cancer Institute, Trinity College Dublin, University of Dublin, Dublin, Ireland.
Background: Prostate bed (PB) motion may lead to geographical miss of the target volume in post-prostatectomy radiotherapy (RT). Optimal clinical target volume (CTV) to planning target volume (PTV) margins prevent geographical miss and unnecessary irradiation of normal tissue. There is little data available informing appropriate CTV to PTV margins in the post-prostatectomy setting.
View Article and Find Full Text PDFRadiat Oncol
November 2024
Radiation Oncology Key Laboratory of Sichuan Province, Sichuan Cancer Hospital & Institute, Sichuan Cancer Center, School of Medicine, University of Electronic Science and Technology of China, No.55, Section 4, South Renmin Road, Chengdu, China.
Background: The aim of this study was to evaluate the inter-fraction reproducibility and intra-fraction stability of breast radiotherapy using voluntary deep-inspiration breath hold (DIBH) and free breathing (FB) based on an optical surface imaging system (OSIS).
Methods: Seventeen patients (510 breath-hold sessions) treated using a field-in-field (FiF) technique and twenty patients (600 breath-free sessions) treated with a volume-modulated arc therapy (VMAT) technique were included in this retrospective study. All the patients were positioned with the guidance of CBCT and OSIS, and also monitored with OSIS throughout the whole treatment session.
Radiother Oncol
December 2024
Cancer Division, University College London Hospitals NHS Foundation Trust, London, UK; Department of Medical Physics and Biomedical Engineering, University College London, London, UK.
J Appl Clin Med Phys
December 2024
Medical Physics Department, Hospital Universitario Puerta de Hierro Majadahonda, Madrid, Spain.
Purpose: The aim of this study was to assess the accuracy of a surface-guided radiotherapy (SGRT) system for setup and intra-fraction motion control in frameless non-coplanar stereotactic radiosurgery (fSRS) using actual patient data immobilized with two different types of open-faced masks and employing a novel SGRT systems settings.
Methods And Materials: Forty-four SRS patients were immobilized with two types of open-faced masks. Sixty lesions were treated, involving the analysis of 68 cone-beam scans (CBCT), 157 megavoltage (MV) images, and 521 SGRT monitoring sessions.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!