We aimed to develop a software for facilitating absorbed dose per pixel (pixel dose) calculation using a size-specific dose estimate (SSDE). We calculated the pixel dose at nine equal points inserted into the radiophotoluminescence glass dosemeter (RPLD) and compared the pixel dose with the measured doses using RPLD. With this method, the relative errors of average pixel dose was -0.1% for adults and 2.86, 3.36 and 1.17% for those aged 10, 5 and 1 years without tube current modulation, respectively. In contrast, the relative error of SSDE was 17.37% for adults and 20.38, 20.73 and 19.20% for those aged 10, 5 and 1 years, respectively. In other words, the pixel dose was almost equal to the measured doses. Therefore, our software can be useful for determining arbitrary point.