A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Comparison of the Biofire FilmArray Respiratory Panel, Seegene AnyplexII RV16, and Argene for the detection of respiratory viruses. | LitMetric

Background: Respiratory infections are common reasons for hospital admission, and are associated with enormous economic burden due to significant morbidity and mortality. The wide spectrum of microbial agents underlying the pathology renders the diagnosis of respiratory infections challenging. Molecular diagnostics offer an advantage to the current serological and culture-based methods in terms of sensitivity, coverage, hands-on time, and time to results.

Objectives: This study aimed to compare the clinical performance of three commercial kits for respiratory viral detection.

Study Design: The performance of FilmArray Respiratory Panel, AnyplexII RV16, and Argene was compared using clinical respiratory samples (n = 224, comprising 189 nasopharyngeal swabs in Universal Transport Medium (UTM) and 35 endotracheal aspirates), based on common overlapping targets across the platforms. Influenza A "equivocal" and "no-subtype" samples by FilmArray were further compared to a laboratory-developed Influenza A/B test.

Results And Conclusions: The overall performance of all three platforms appeared to be comparable with regards to sensitivities (95.8-97.9%) and specificities (96.1-98.0%), detection of coinfections, and distinguishment of influenza from non-influenza cases. "Equivocal" and "no-subtype" samples by FilmArray mostly represented weak Influenza A by laboratory-developed test. Lower respiratory tract samples had comparable final-run success-rates and discordant-rates as compared to UTM. Coronavirus HKU1, which was not targeted by AnyplexII RV16, were detected as OC43. The expected test volume would be the main determinant for the selection of platform. Among the platforms, the FilmArray is the most automated but is of the lowest-throughput and has the highest reagent cost.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7185839PMC
http://dx.doi.org/10.1016/j.jcv.2018.07.002DOI Listing

Publication Analysis

Top Keywords

anyplexii rv16
12
respiratory
8
filmarray respiratory
8
respiratory panel
8
rv16 argene
8
respiratory infections
8
performance three
8
"equivocal" "no-subtype"
8
"no-subtype" samples
8
samples filmarray
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!