Objective: This study aims to compare the informative value of a capacitively coupled electrocardiogram (cECG) to a conventional galvanic reference ECG (rECG) in patients after a major cardiac event under simulated driving conditions. Addressed research questions are the comparison and coherence of cECG and rECG by means of the signal quality, the artifact rate, the rate of assessable data for differential diagnosis, the visibility of characteristic ECG structures in cECG, the precision of ECG time intervals, and heart rate (in particular, despite possible waveform deformations due to the cardiac preconditions).

Methods: In a clinical trial, cECG and rECG data were recorded from ten patients after a major cardiac event. The cECG and rECG data were blindly evaluated by two cardiologists with regard to signal quality, artifacts, assessable data for differential diagnosis, visibility of ECG structures, and ECG time intervals. The results were statistically compared.

Results: The cECG presented with more artifacts, an inferior signal quality, and less assessable data. However, when the data were assessable, determination of the ECG interval lengths was coherent to the one obtained from the rECG.

Conclusion: When the signal quality is sufficient, the cECG yields the same informative value as the rECG.

Significance: For certain scenarios, cECG might replace rECG systems. Hence, it is an important research question whether a similar amount of information can be obtained using a cECG system.

Download full-text PDF

Source
http://dx.doi.org/10.1109/TBME.2018.2855661DOI Listing

Publication Analysis

Top Keywords

signal quality
16
cecg recg
12
assessable data
12
cecg
9
simulated driving
8
patients major
8
major cardiac
8
cardiac event
8
data differential
8
differential diagnosis
8

Similar Publications

Liquid biopsy technologies: innovations and future directions in breast cancer biomarker detection.

Biomed Microdevices

January 2025

Department of Laboratory Medicine, The Affiliated Hospital of Southwest Medical University, 25 Taiping Street, Luzhou, 646000, Sichuan, People's Republic of China.

Globally, breast cancer is the most frequent type of cancer, and its early diagnosis and screening can significantly improve the probability of survival and quality of life of those affected. Liquid biopsy-based targets such as circulating tumor cells, circulating tumor DNA, and exosomes have been instrumental in the early discovery of cancer, and have been found to be effective in stage therapy, recurrence monitoring, and drug selection. Biosensors based on these target related biomarkers convert the tested substances into quantifiable signals such as electrical and optical signals through signal transduction, which has the advantages of high sensitivity, simple operation, and low invasiveness.

View Article and Find Full Text PDF

Purpose: To evaluate the effects of four-dimensional noise reduction filtering using a similarity algorithm (4D-SF) on the image quality and tumor visibility of low-dose dynamic computed tomography (CT) in evaluating breast cancer.

Materials And Methods: Thirty-four patients with 38 lesions who underwent low-dose dynamic breast CT and were pathologically diagnosed with breast cancer were enrolled. Dynamic CT images were reconstructed using iterative reconstruction alone or in combination with 4D-SF.

View Article and Find Full Text PDF

Mitochondria, the cellular powerhouses, are pivotal to neuronal function and health, particularly through their role in regulating synaptic structure and function. Spine reprogramming, which underlies synapse development, depends heavily on mitochondrial dynamics-such as biogenesis, fission, fusion, and mitophagy as well as functions including ATP production, calcium (Ca) regulation, and retrograde signaling. Mitochondria supply the energy necessary for assisting synapse development and plasticity, while also regulating intracellular Ca homeostasis to prevent excitotoxicity and support synaptic neurotransmission.

View Article and Find Full Text PDF

Cavity as radio telescope for galactic dark photon.

Sci Bull (Beijing)

January 2025

School of Physics and State Key Laboratory of Nuclear Physics and Technology, Peking University, Beijing 100871, China; Center for High Energy Physics, Peking University, Beijing 100871, China; Key Laboratory of Particle Acceleration Physics and Technology, Chinese Academy of Sciences, Beijing 100049, China. Electronic address:

Dark photons, as a minimal extension of the Standard Model through an additional Abelian gauge group, may propagate relativistically across the galaxy, originating from dark matter decay or annihilation, thereby contributing to a galactic dark photon background. The generation of dark photons typically favors certain polarization modes, which are dependent on the interactions between dark matter and dark photons. We introduce a framework in which a resonant cavity is utilized to detect and differentiate these polarizations, leveraging the daily variation in expected signals due to the anisotropic distribution of dark photons and the rotation of the Earth.

View Article and Find Full Text PDF

Risk of bias assessment tools often addressed items not related to risk of bias and used numerical scores.

J Clin Epidemiol

January 2025

Division of Nephrology and Hypertension, Department of Internal Medicine, University of Kansas Medical Centre, 3901 Rainbow Blvd, MS3002, Kansas City, KS, USA. Electronic address:

Objective: We aimed to determine whether the existing risk of bias assessment tools addressed constructs other than risk of bias or internal validity, and whether they used numerical scores to express quality, which is discouraged and may be a misleading approach.

Methods: We searched Ovid MEDLINE and Embase to identify quality appraisal tools across all disciplines in human health research. Tools designed specifically to evaluate reporting quality were excluded.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!