Background: Over the last decade, cuffed endotracheal tubes are increasingly used in pediatric anesthesia and also in pediatric intensive care. However, the smaller inner diameter of cuffed endotracheal tubes and, implicitly, the increased endotracheal tube resistance is still a matter of debate.

Aims: This in vitro study investigated work of breathing and inspiratory airway pressures in cuffed and uncuffed endotracheal tubes and the impact of pressure support ventilation and automatic tube compensation.

Methods: In 5 simulated neonatal and pediatric lung models, the Active Servo Lung 5000 and an intensive care ventilator were used to quantify the differences in work of breathing under spontaneous breathing (with and without pressure support ventilation and automatic tube compensation) between cuffed and uncuffed endotracheal tubes. Additionally, differences in inspiratory airway pressures, measured either proximal or distal of the endotracheal tube, between cuffed and uncuffed endotracheal tubes under mechanical ventilation were investigated.

Results: Work of breathing was overall 10.27% [95% confidence interval 9.01-11.94] higher with cuffed than with uncuffed endotracheal tubes and was dramatically reduced by 34.19% [95% confidence interval 31.61-35.25] with the application of pressure support. Automatic tube compensation almost diminished work of breathing differences between the 2 endotracheal tube types in nearly all pediatric lung models. Peak inspiratory and mean airway pressures measured at the proximal endotracheal tube end revealed significantly higher values in cuffed than in uncuffed endotracheal tubes. However, these differences measured at the distal end of the endotracheal tube became minimal.

Conclusion: This in vitro study confirmed significant differences in work of breathing and inspiratory pressures between cuffed and uncuffed endotracheal tubes. Work of breathing, however, is almost neutralized by pressure support ventilation with automatic tube compensation and distal inspiratory airway pressures that, from a clinical perspective, are not significantly increased.

Download full-text PDF

Source
http://dx.doi.org/10.1111/pan.13430DOI Listing

Publication Analysis

Top Keywords

endotracheal tubes
36
work breathing
28
cuffed uncuffed
28
uncuffed endotracheal
24
endotracheal tube
20
inspiratory airway
16
airway pressures
16
pressure support
16
automatic tube
16
endotracheal
14

Similar Publications

Fungal infections related to biofilm formation on medical devices, such as endotracheal tubes (ETTs), pose significant health risks, especially during intubation procedures where fungi like Candida spp. can migrate into the lower respiratory tract. This study explores the use of Photodynamic Therapy (PDT) to prevent fungal cell migration from ETT surfaces to lungs, focusing on the role of curcumin as a photosensitizer.

View Article and Find Full Text PDF

Management of the airway in austere environments can differ substantially from standard in-hospital airway management. Devices such as nasopharyngeal airways, oropharyngeal airways, endotracheal tubes, extraglottic airways, ventilators, and sedative and paralytic medications may not be available. Weather, scene hazards, difficulties of extrication, transport times, skill sets of rescuers, and availability of advanced equipment are highly variable.

View Article and Find Full Text PDF

Objectives: Idiopathic subglottic stenosis(ISS) is an airway stricture between larynx and trachea, within 1 cm from the vocal cords. Resection-reconstruction presents technical issues because of the need to resect the anterior portion of the cricoid cartilage, but not the posterior plate beyond which the recurrent laryngeal nerves access the vocal cords. The main surgical challenge consists of the minute airway caliber short below the cords, ensuing after the resection.

View Article and Find Full Text PDF

Objective: To assess whether training provided to an inexperienced clinician just before performing a high stakes procedure can improve procedural care quality, measuring the first attempt success rate of trainees performing infant orotracheal intubation.

Design: Randomized clinical trial.

Setting: Single center, quaternary children's hospital in Boston, MA, USA.

View Article and Find Full Text PDF

Background: Sore throat and hoarseness are two common complications of intubation in patients with general anesthesia. This research aimed to compare the effect of different endotracheal cuff pressures on sore throat and hoarseness after general anesthesia.

Materials And Methods: The present double-blinded clinical trial was conducted on 45 patients who are candidates for surgery with general anesthesia in autumn and winter 2021.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!