Clinical Feasibility of Single-Source Dual-spiral 4D Dual-Energy CT for Proton Treatment Planning Within the Thoracic Region.

Int J Radiat Oncol Biol Phys

OncoRay - National Center for Radiation Research in Oncology, Faculty of Medicine and University Hospital Carl Gustav Carus, Technische Universität Dresden, Helmholtz-Zentrum Dresden-Rossendorf, Dresden, Germany; Helmholtz-Zentrum Dresden-Rossendorf, Institute of Radiooncology - OncoRay, Dresden, Germany; Department of Radiotherapy and Radiation Oncology, Faculty of Medicine and University Hospital Carl Gustav Carus, Technische Universität Dresden, Dresden, Germany.

Published: November 2018

Purpose: Single-source dual-spiral dual-energy computed tomography (DECT) provides additional patient information but is prone to motion between the 2 consecutively acquired computed tomography (CT) scans. Here, the clinical applicability of dual-spiral time-resolved DECT (4D-DECT) for proton treatment planning within the thoracic region was evaluated.

Methods And Materials: Dual-spiral 4D-DECT scans of 3 patients with lung cancer were acquired. For time-averaged datasets and 4 breathing phases, the geometric conformity of 80 kVp and 140 kVp 4D-DECT scans before image post-processing was assessed by normalized cross correlation (NCC). Additionally, the conformity of the corresponding DECT-derived 58 keV and 79 keV pseudo-monoenergetic CT datasets after image post-processing, including deformable image registration (DIR), was determined. To analyze the reliability of proton dose calculation, clinical (Plan) and artificial worst-case (Plan, targeting the diaphragm) treatment plans were calculated on 140 kVp and 79 keV datasets and compared with gamma analyses (0.1% dose-difference and 1 mm distance-to-agreement criterion). The applicability of a patient-specific DECT-based prediction of stopping-power ratio (SPR) was investigated and proton range shifts compared with the clinical heuristic CT-number-to-SPR conversion were assessed. Finally, the delineation variability of an experienced radiation oncologist was quantified.

Results: Dual-spiral 4D-DECT scans without DIR showed a high geometric conformity, with an average NCC ± standard deviation of 98.7% ± 1.0% when including all patient voxels or 88.2% ± 7.8% when considering only lung. DIR improved the conformity, leading to an average NCC of 99.9% ± 0.1% and 99.6% ± 0.5%, respectively. Plan dose distributions on 140 kVp and 79 keV datasets were similar, with an average gamma passing rate of 99.9% (99.2%-100%). The worst-case evaluation still revealed high passing rates (99.3% on average, 92.4% as minimum). Clinically relevant mean range shifts of 2.2% ± 1.2% were determined between patient-specific DECT-based SPR prediction and clinical heuristic CT-number-to-SPR conversion. The intra-observer delineation variability was slightly reduced using additional DECT-derived datasets.

Conclusions: The 79 keV pseudo-monoenergetic CT datasets can be consistently obtained from dual-spiral 4D-DECT and are applicable for dose calculation. Patient-specific DECT-based SPR prediction performed well and potentially reduces range uncertainty in proton therapy of patients with lung cancer.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ijrobp.2018.06.044DOI Listing

Publication Analysis

Top Keywords

dual-spiral 4d-dect
12
4d-dect scans
12
140 kvp
12
patient-specific dect-based
12
single-source dual-spiral
8
dual-spiral dual-energy
8
proton treatment
8
treatment planning
8
planning thoracic
8
thoracic region
8

Similar Publications

Clinical Feasibility of Single-Source Dual-spiral 4D Dual-Energy CT for Proton Treatment Planning Within the Thoracic Region.

Int J Radiat Oncol Biol Phys

November 2018

OncoRay - National Center for Radiation Research in Oncology, Faculty of Medicine and University Hospital Carl Gustav Carus, Technische Universität Dresden, Helmholtz-Zentrum Dresden-Rossendorf, Dresden, Germany; Helmholtz-Zentrum Dresden-Rossendorf, Institute of Radiooncology - OncoRay, Dresden, Germany; Department of Radiotherapy and Radiation Oncology, Faculty of Medicine and University Hospital Carl Gustav Carus, Technische Universität Dresden, Dresden, Germany.

Purpose: Single-source dual-spiral dual-energy computed tomography (DECT) provides additional patient information but is prone to motion between the 2 consecutively acquired computed tomography (CT) scans. Here, the clinical applicability of dual-spiral time-resolved DECT (4D-DECT) for proton treatment planning within the thoracic region was evaluated.

Methods And Materials: Dual-spiral 4D-DECT scans of 3 patients with lung cancer were acquired.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!