Prototypes of a PZT-based ultrasound needle-actuating device have shown the ability to reduce needle penetration force and enhance needle visibility with color Doppler imaging during needle insertion for tissue biopsy and regional anesthesia. However, the demand for smaller, lighter devices and the need for high performance transducers have motivated investigation of a different configuration of needle-actuation transducer, utilizing the -mode of PZT4 piezoceramic, and exploration of further improvement in its performance using relaxor-type piezocrystal. This paper outlines the development of the -mode needle actuation transducer design from simulation to fabrication and demonstration. Full characterization was performed on transducers for performance comparison. The performance of the proposed smaller, lighter -mode transducer is comparable with that of previous -mode transducers. Furthermore, it has been found to be much more efficient when using PMN-PT piezocrystal rather than piezoceramic.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1109/TUFFC.2018.2838566 | DOI Listing |
Ultrasonics
March 2024
Centre for Medical & Industrial Ultrasonics, James Watt School of Engineering, University of Glasgow, Glasgow G12 8QQ, UK. Electronic address:
Mn:PIN-PMN-PT piezocrystal is investigated to determine whether its enhanced energy density makes it a candidate transducer material for power ultrasonics applications. To this end, the electromechanical and vibrational characteristics of a simple configuration of a bolted Langevin transducer (BLT) and then an ultrasonic surgical device, both incorporating Mn:PIN-PMN-PT piezocrystal, are compared with the same transducer configurations incorporating a conventional hard PZT piezoceramic commonly used in high-power ultrasonic transducers. The material properties of Mn:PIN-PMN-PT are determined using a single sample characterisation technique and these are used in finite element analysis (FEA) to design and then fabricate the BLT and ultrasonic surgical device, tuned to the first and second longitudinal modes at 20 kHz respectively.
View Article and Find Full Text PDFIEEE Trans Ultrason Ferroelectr Freq Control
August 2018
Prototypes of a PZT-based ultrasound needle-actuating device have shown the ability to reduce needle penetration force and enhance needle visibility with color Doppler imaging during needle insertion for tissue biopsy and regional anesthesia. However, the demand for smaller, lighter devices and the need for high performance transducers have motivated investigation of a different configuration of needle-actuation transducer, utilizing the -mode of PZT4 piezoceramic, and exploration of further improvement in its performance using relaxor-type piezocrystal. This paper outlines the development of the -mode needle actuation transducer design from simulation to fabrication and demonstration.
View Article and Find Full Text PDFUltrasonics
July 2016
School of Engineering, University of Glasgow, Glasgow, UK.
The higher performance of relaxor-based piezocrystals compared with piezoceramics is now well established, notably including improved gain-bandwidth product, and these materials have been adopted widely for biomedical ultrasound imaging. However, their use in other applications, for example as a source of focused ultrasound for targeted drug delivery, is hindered in several ways. One of the issues, which we consider here, is in shaping the material into the spherical geometries used widely in focused ultrasound.
View Article and Find Full Text PDFMaterials (Basel)
December 2015
School of Engineering, University of Glasgow, Glasgow G12 8QQ, UK.
Piezocrystals, especially the relaxor-based ferroelectric crystals, have been subject to intense investigation and development within the past three decades, motivated by the performance advantages offered by their ultrahigh piezoelectric coefficients and higher electromechanical coupling coefficients than piezoceramics. Structural anisotropy of piezocrystals also provides opportunities for devices to operate in novel vibration modes, such as the face shear mode, with domain engineering and special crystal cuts. These piezocrystal characteristics contribute to their potential usage in a wide range of low- and high-power ultrasound applications.
View Article and Find Full Text PDFIEEE Trans Ultrason Ferroelectr Freq Control
November 2015
An ultrasonic microcutter is an alternative approach to conventional ultrasonic instruments actuated by sandwich piezoelectric transducers for surgery. This paper reports high-power behavior of a microcutter actuated by the piezocrystal lead magnesium niobate-lead titanate (PMN-PT), defining its practical performance and the feasibility of PMNPT actuation for surgical applications. The microcutter was driven at resonance with constant current amplitudes, either unloaded or loaded by poultry breast tissue, until its behavior achieved a steady state.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!