Goal: An accurate rhythm analysis during cardiopulmonary resuscitation (CPR) would contribute to increase the survival from out-of-hospital cardiac arrest. Piston-driven mechanical compression devices are frequently used to deliver CPR. The objective of this paper was to design a method to accurately diagnose the rhythm during compressions delivered by a piston-driven device.
Methods: Data was gathered from 230 out-of-hospital cardiac arrest patients treated with the LUCAS 2 mechanical CPR device. The dataset comprised 201 shockable and 844 nonshockable ECG segments, whereof 270 were asystole (AS) and 574 organized rhythm (OR). A multistage algorithm (MSA) was designed, which included two artifact filters based on a recursive least squares algorithm, a rhythm analysis algorithm from a commercial defibrillator, and an ECG-slope-based rhythm classifier. Data was partitioned randomly and patient-wise into training (60%) and test (40%) for optimization and validation, and statistically meaningful results were obtained repeating the process 500 times.
Results: The mean (standard deviation) sensitivity (SE) for shockable rhythms, specificity (SP) for nonshockable rhythms, and the total accuracy of the MSA solution were: 91.7 (6.0), 98.1 (1.1), and 96.9 (0.9), respectively. The SP for AS and OR were 98.0 (1.7) and 98.1 (1.4), respectively.
Conclusions: The SE/SP were above the 90%/95% values recommended by the American Heart Association for shockable and nonshockable rhythms other than sinus rhythm, respectively.
Significance: It is possible to accurately diagnose the rhythm during mechanical chest compressions and the results considerably improve those obtained by previous algorithms.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1109/TBME.2018.2827304 | DOI Listing |
J Sleep Res
January 2025
Department of Neurology, University of Washington School of Medicine, Seattle, Washington, USA.
Changes in social zeitgebers across the lifespan affect the interaction between biological and social clocks, potentially contributing to social jetlag. Extant literature suggests a reduction in social jetlag given declining social obligations occurring after retirement, but is limited to self-reported methods and cross-sectional designs. Leveraging longitudinal and ecologically valid data from consumer sleep technology, we analysed objective sleep data from 2439 users of the polysomnography-validated SleepScore mobile application, encompassing 500,415 total nights recorded.
View Article and Find Full Text PDFBMC Med Genomics
January 2025
Department of Otolaryngology, First Affiliated Hospital of Kunming Medical University, 295 Xichang Road, WuHua District, Kunming City, Yunnan Province, China.
Hearing loss is a prevalent condition with a significant impact on individuals' quality of life. However, comprehensive studies investigating the differential gene expression and regulatory mechanisms associated with hearing loss are lacking, particularly in the context of diverse patient samples. In this study, we integrated data from 10 patients across different regions, age groups, and genders, with their data retrieved from a public transcriptome database, to explore the molecular basis of hearing loss.
View Article and Find Full Text PDFInt J Cardiol
January 2025
Department of Cardiology, University Hospital Královské Vinohrady, Prague, Czech Republic; Third Faculty of Medicine, Charles University, Prague, Czech Republic.
Background And Aims: Myocardial infarction (MI) in multivessel disease (MVD) and chronic total occlusion (CTO) is associated with high mortality. However, all-cause mortality of matched cohort without a CTO is unclear. Our aim was to analyse clinical characteristics, presenting symptoms, and survival of patients with MI in MVD and the possible impact of CTO on 1-year mortality.
View Article and Find Full Text PDFHeliyon
January 2025
Faculty of Medicine and Medical Sciences, University of Balamand, Kalhat, Lebanon.
Background: Transvenous pacemakers (TVP) and leadless pacemakers (LP) are two reliable permanent modalities for the treatment of heart rhythm disorders. Several observational studies explored the safety and efficacy of the two devices. The aim of this meta-analysis study is to present a comparative analysis of the safety of leadless versus transvenous pacemakers.
View Article and Find Full Text PDFSci Rep
January 2025
Laboratory for Radiophysical and Optical Methods of Environmental Research, National Research Tomsk State University, Tomsk, Russia, 634050.
Monitoring the parameters and behavior of plankton makes it possible to assess the state of the aquatic ecosystem and detect the beginning of an environmental disaster at an early stage. In this respect, the most informative method for the in situ plankton study is underwater digital holography. This method allows obtaining information on the size, shape, and location of plankton individuals, as well as performing their classification and biotesting according to their behavioral responses using a submersible holographic camera non-invasively, in real time, and in the automatic mode.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!