Purpose: To assess the reliability of a prototype automated supplying artery tracking software (ASATS) using multidetector-row CT (MDCT) images in emergent TAE.

Materials And Methods: Consecutive 53 patients underwent 57 sessions of emergent TAE during 7 months. Twenty-one cases were excluded due to a lack of CT data (n = 12) or negative angiographic findings (n = 9). Remaining 34 sessions of TAE and MDCT images in 32 patients (mean age 62.9 years; age range 37-92 years) were enrolled. ASATS was retrospectively conducted for the identification of supplying arteries which were confirmed with angiography (automated method). Manual modification was added as needed (semi-automated method). Two observers independently reviewed the MDCT images to detect supplying arteries (manual method). Detectability of supplying artery and time to analysis were compared among the automated, semi-automated, and manual methods by both observers.

Results: A total of 64 bleeding sites were demonstrated on angiography. The detectability was 28 (43.8%) for automated method, 53 (82.8%) for semi-automated method, 55 (85.9%) for observer 1, and 58 (90.6%) for observer 2. Detectability of semi-automated method was significantly better than of automated method (P = 0.000) and comparable with manual method by both observers (P = 0.193 and 0.081). Average time to analysis was 185.4 s for automated method, 297.2 s for semi-automated method, 186.2 s for observer 1, and 243.7 s for observer 2.

Conclusion: ASATS has a sufficient ability to identify supplying arteries of bleeding by adding manual modification as needed and can be used for emergent TAE.

Level Of Evidence: Level 4, Case Control Study.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00270-018-2027-8DOI Listing

Publication Analysis

Top Keywords

automated method
16
semi-automated method
16
supplying artery
12
mdct images
12
supplying arteries
12
method
10
automated supplying
8
artery tracking
8
tracking software
8
images emergent
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!