  • We utilized a shifted-excitation Raman difference spectroscopy method to classify breast tissue samples (normal, fibroadenoma, invasive carcinoma) with 240 measurements per tissue type.
  • Although raw spectra provided clear differentiation between normal and tumor tissues, distinguishing fibroadenoma from invasive carcinoma required isolating pure Raman spectra due to fluorescence interference.
  • The method achieved high diagnostic accuracy, especially for invasive carcinoma (99.15% sensitivity) and overall demonstrated potential as a support tool for traditional histopathological diagnosis in breast pathology.

Article Abstract

We used a shifted-excitation Raman difference spectroscopy method for the ex vivo classification of resected and formalin-fixed breast tissue samples as normal (healthy) tissue, fibroadenoma, or invasive carcinoma. We analyzed 8 tissue samples containing invasive carcinoma that were surrounded by normal tissue and 3 tissue samples with fibroadenoma only. We made various measurement sites on various tissue samples, in total 240 measurements for each type of tissue. Although the acquired raw spectra contain enough information to clearly differentiate between normal and tumor (fibroadenoma and invasive carcinoma) tissue, the differentiation between fibroadenoma and invasive carcinoma was possible only after the shifted-excitation Raman difference spectroscopy isolation of pure Raman spectra from the heavily fluorescence interfered raw spectra. We used 784 and 785 nm as excitation wavelengths for the shifted-excitation Raman difference spectroscopy method. The differences in the obtained pure Raman spectra are assigned to the different chemical compositions of normal breast tissue, fibroadenoma, and invasive breast carcinoma. Principal component analysis and linear discriminant analysis showed excellent classification results in the Raman shift range between 1000 and 1800 cm. Invasive breast carcinoma was identified with 99.15% sensitivity, and the absence of invasive carcinoma was identified with 90.40% specificity. Tumor tissue in tumor-containing tissue was identified with 100% sensitivity, and the absence of tumor in no-tumor containing tissue was identified with 100% specificity. As gold standard for the determination of the sensitivity and the specificity, we considered the conventional histopathological classification. In summary, shifted-excitation Raman difference spectroscopy could be potentially very useful to support histopathological diagnosis in breast pathology.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6048663PMC
http://dx.doi.org/10.1177/1533033818782532DOI Listing

Publication Analysis

Top Keywords

shifted-excitation raman
20
raman difference
20
difference spectroscopy
20
invasive carcinoma
20
tissue samples
16
fibroadenoma invasive
16
tissue
12
raman
8
spectroscopy method
8
breast tissue
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!