Dosimetry in a mammography phantom using TLD-300 dosimeters.

Med Phys

Instituto de Física, Universidad Nacional Autónoma de México, 04510, Mexico City, Mexico.

Published: July 2018

Purpose: The purpose of this study was to evaluate the photon field effective energy (E ) distribution and percentage depth-dose (PDD) within a mammography phantom by the analysis of the CaF :Tm (TLD-300) thermoluminescent (TL) glow curve. The experimental procedure involves the use of TLD-300 to determine with single dosimeter exposures both the relative dose and the beam quality.

Methods: TLD-300 chips were exposed to x rays from a GE Senographe 2000D mammography unit at the surface and different depths within a BR12 phantom. X-ray beams were generated with Mo/Mo, Mo/Rh, and Rh/Rh anode/filter combinations and voltages between 25 and 34 kV. Glow curves were deconvoluted into component peaks and the high- to low-temperature ratio (HLTR) was evaluated. The photon field E was obtained from the HLTR values using a calibration curve determined previously. PDD was established from the peak 5 TL signal (TLS ) at depths between 0.0 and 3.5 cm inside the phantom. Taking into account the differences in density and composition between CaF :Tm and breast tissue, an energy-dependent correction factor (β(E)) was applied to TLS . Measurements were compared with radiation transport Monte Carlo (MC) simulations performed with PENELOPE-2008.

Results: A typical 5% change in the HLTR from the phantom top surface to 3.5 cm depth was measured, which corresponds to a 2.2 keV increase in photon field E . Values of the β(E) correction factor were 0.33 and 0.13 for E equal to 15.1 and 22.5 keV, respectively. This strong energy dependence of β(E) is mostly due to the differences in fluence attenuation between CaF and breast tissue. According to PDD measurements, dose decreased to half the surface value at depths between 0.7 and 1.0 cm for Mo/Mo/25 and Rh/Rh/34 beams, respectively. Values of PDD, less than 10% at 3.5 cm depth, would have been overestimated by about 3.5% (a large relative error) if an energy-independent correction factor had been assumed. Mean differences between experiments and MC simulations were 0.8 keV and 1.2% in the determination of E and PDD, respectively.

Conclusion: The TLD-300 glow curve was used to accurately measure the photon field E and PDD within a mammographic phantom. This work has demonstrated that E and dose can be established simultaneously by using solely TLD-300.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.13084DOI Listing

Publication Analysis

Top Keywords

photon field
16
correction factor
12
mammography phantom
8
glow curve
8
surface depths
8
caf breast
8
breast tissue
8
35 cm depth
8
phantom
6
tld-300
6

Similar Publications

Virtual staining from bright-field microscopy for label-free quantitative analysis of plant cell structures.

Plant Mol Biol

January 2025

Graduate School of Science and Technology, Kumamoto University, 2-39-1 Kurokami, Chuo-Ku, Kumamoto, 860-8555, Japan.

The applicability of a deep learning model for the virtual staining of plant cell structures using bright-field microscopy was investigated. The training dataset consisted of microscopy images of tobacco BY-2 cells with the plasma membrane stained with the fluorescent dye PlasMem Bright Green and the cell nucleus labeled with Histone-red fluorescent protein. The trained models successfully detected the expansion of cell nuclei upon aphidicolin treatment and a decrease in the cell aspect ratio upon propyzamide treatment, demonstrating its utility in cell morphometry.

View Article and Find Full Text PDF

Near-infrared (NIR) spectroscopy with chemometrics was used to evaluate semi-refined flaxseed protein extract (SRFPE) and predict its protein content. The effect of extraction medium (water vs. alkaline) employed after pulsed electric field (PEF) treatment on protein yield was also studied.

View Article and Find Full Text PDF

Proximity-Induced Superconductivity in Ferromagnetic FeGeTe and Josephson Tunneling through a van der Waals Heterojunction.

ACS Nano

January 2025

International Center for Quantum Design of Functional Materials (ICQD), Hefei National Laboratory for Physical Sciences at Microscale, University of Science and Technology of China, Hefei 230026, China.

Synergy between superconductivity and ferromagnetism may offer great opportunities in nondissipative spintronics and topological quantum computing. Yet at the microscopic level, the exchange splitting of the electronic states responsible for ferromagnetism is inherently incompatible with the spin-singlet nature of conventional superconducting Cooper pairs. Here, we exploit the recently discovered van der Waals ferromagnets as enabling platforms with marvelous controllability to unravel the myth between ferromagnetism and superconductivity.

View Article and Find Full Text PDF

A review of state-of-the-art resolution improvement techniques in SPECT imaging.

EJNMMI Phys

January 2025

Department of Nuclear Medicine, Division of Life Sciences and Medicine, The First Affiliated Hospital of USTC, University of Science and Technology of China, Hefei, 230001, Anhui, China.

Single photon emission computed tomography (SPECT), a technique capable of capturing functional and molecular information, has been widely adopted in theranostics applications across various fields, including cardiology, neurology, and oncology. The spatial resolution of SPECT imaging is relatively poor, which poses a significant limitation, especially the visualization of small lesions. The main factors affecting the limited spatial resolution of SPECT include projection sampling techniques, hardware and software.

View Article and Find Full Text PDF

Despite the outstanding progress in photonic sensor devices, a major limitation for its application as label-free biosensors for biomedical analysis lies in the surface biofunctionalization step, that is, the reliable immobilization of the biorecognition element onto the sensor surface. Here, we report the integration of bottom-up synthesized nanoporous graphene onto bimodal waveguide interferometric biosensors as an atomically precise biofunctionalization scaffold. This combination leverages the high sensitivity of bimodal waveguide interferometers and the large functional surface area of nanoporous graphene to create highly sensitive, selective, and robust biosensors for the direct immunoassay detection of C-reactive protein (CRP), an inflammatory biomarker widely used in the clinical diagnosis of infections and sepsis.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!