Purpose: To evaluate the effectiveness of Spot photoscreener and SW800 vision screener in detecting amblyopia risk factors in Chinese children between 4 and 6 years of age.

Methods: One hundred and thirteen children (226 eyes) underwent complete ophthalmologic examination, cycloplegic retinoscopy refraction, prism cover tests and photoscreen using both Spot (v2.1.4) and SW800 (v1.0.1.0) photoscreeners. The agreement of results obtained from photoscreener and retinoscopy was evaluated by paired t-test as well as Pearson correlation test. The sensitivity and specificity of detecting amblyopia risk factors were calculated based on the American Association of Pediatric Ophthalmology and Strabismus 2013 guidelines. The overall effectiveness of detecting amblyopia risk factors by using either photoscreener was analysed by receiver operating characteristic (ROC) curves.

Results: A strong linear agreement was observed between Spot and retinoscopy (p<0.01) in aspects of spherical equivalent (SE, Pearson's r=0.95), dioptre sphere (DS, r=0.97), dioptre cylinder (DC, r=0.84) and horizontal deviation (Hdev, r=0.91), with overall -0.17 D myopic shift of SE. Significant correlation was also shown between SW800 and retinoscopy (p<0.01) in aspects of SE (r=0.90), DS (r=0.93), DC (r=0.82) and Hdev (r=0.80), with overall -0.12 D myopic shift of SE. The overall sensitivity and specificity in detecting amblyopia risk factors were 94.0% and 80.0% for Spot and 88.8% and 81.1% for SW800.

Conclusion: The measurements of Spot and SW800 photoscreener showed a strong agreement with cycloplegic retinoscopy refraction and prism cover tests. The performance of both screeners in detecting individual amblyopia risk factors is satisfactory. ROC analysis indicates that the Spot and SW800 performed very similarly in detecting amblyopia risk factors.

Download full-text PDF

Source
http://dx.doi.org/10.1136/bjophthalmol-2018-311885DOI Listing

Publication Analysis

Top Keywords

detecting amblyopia
12
amblyopia risk
12
risk factors
12
chinese children
8
compared performance
4
spot
4
performance spot
4
spot sw800
4
sw800 photoscreeners
4
photoscreeners chinese
4

Similar Publications

Introduction: Amblyopia, the leading cause of monocular childhood vision loss, affects millions and is projected to increase. Early detection and treatment are crucial for preventing vision impairment.

Areas Covered: This commentary reviews the current state and opportunities for improvement in amblyopia screening strategy and technology focused primarily within the United States.

View Article and Find Full Text PDF

Playing stereoscopic video games enhances the precision but not the accuracy of depth perception.

Sci Rep

December 2024

Herbert Wertheim School of Optometry and Vision Science, University of California, Berkeley, CA, 94720, USA.

Playing three-dimensional (3D) video games enhances stereo acuity (i.e., the precision of stereopsis) in young adults with normal vision Here we asked whether the improvement in stereoacuity was the result of a reduced disparity pedestal, as indicated by a decreased subjective depth bias (i.

View Article and Find Full Text PDF

Introduction: The prevalence of myopia has increased significantly in recent years including an earlier onset of myopia development on the pediatric population. The main objective of the study is to compare CUVAF (Conjunctival Ultraviolet Autofluorescence) in children with and without myopia to validate its usefulness as an outdoor protective biomarker.

Methods: A case-control observational study was conducted in a child cohort from subjects that attended to the Ophthalmology Department of Clínica Universidad de Navarra for an ophthalmological examination.

View Article and Find Full Text PDF

Modeling monocular form deprivation in rabbits using a simulated-cataract intraocular lens.

Int J Ophthalmol

December 2024

Eye Hospital and School of Ophthalmology and Optometry, Wenzhou Medical University, Wenzhou 325003, Zhejiang Province, China.

Aim: To establish an animal model of form deprivation amblyopia based on a simulated cataract intraocular lens (IOLs).

Methods: Poly(dimethyl siloxane)-SiO thin films (PSF) with different degrees of opacity as IOL materials were prepared. The light transmission of the PSF-IOL was measured, and its biosafety was determined by cell counting kit (CCK)-8 assay using the HLEC-B3 cell line and ARPE-19 cell line.

View Article and Find Full Text PDF

Factors Influencing Refractive Surgery Rejection in Israeli Patients.

Isr Med Assoc J

December 2024

Enaim Refractive Surgery Center, Jerusalem, Israel, Department of Ophthalmology, Tel Aviv Sourasky Medical Center, Tel Aviv, Israel, Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel.

Background: One of the major causes of reversible visual impairment is a refractive error, which can be corrected through refractive surgery. Data regarding the outcomes and complications of these procedures exist; however, there is a notable gap in understanding the factors leading to patient rejection, particularly in diverse populations like Israel.

Objectives: To detect clinical risk factors of patients who intend to undergo LASIK procedure and to improve the conversion rates of LASIK procedure in this population.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!