Purpose: Polychromatic x-rays are used in most computed tomography scanners. In this case, a beam-hardening effect occurs, which degrades the image quality and distorts the shapes of objects in the reconstructed images. When the beam-hardening artifact is not severe, conventional correction methods can reduce the artifact reasonably well. However, highly dense materials, such as iron and titanium, can produce more severe beam-hardening artifacts, which often cannot be corrected by conventional methods. Moreover, when the size of the metal is large, severe darks bands due to photon starvation as well as beam-hardening are generated. The purpose of our study was to develop a new method for correcting severe beam-hardening artifacts and severe dark bands using a high-order polynomial correction function and a prior-image-based linearization method.

Methods: The initial estimate of an image free of beam-hardening (a prior image) was constructed from the initial reconstruction of the original projection data. Its corresponding beam-hardening-free projection data (a prior projection) were calculated by a projection operator onto the prior image. A new beam-hardening correction function G(p ) with many high-order terms was effectively determined via a simple minimization process applied to the difference between the original projection data and the prior projection data. Using the determined correction function G(p ), a corrected linearized sinogram p can be obtained, which became effectively linear for the line integrals of the object. Final beam-hardening corrected images can be reconstructed from the linearized sinogram. The proposed method was evaluated in both simulation and real experimental studies.

Results: All investigated cases in both simulations and real experiments showed that the proposed method effectively removed not only streaks for moderate beam-hardening artifacts but also dark bands for severe beam-hardening artifacts without causing structural and contrast distortion.

Conclusions: The prior-image-based linearization method exhibited better correction performance than conventional methods. Because the proposed method did not require time-consuming iterative reconstruction processes to obtain the optimal correction function, it can expedite the correction procedure and incorporate more high-order terms in the linearization correction function in comparison to the conventional methods.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.13072DOI Listing

Publication Analysis

Top Keywords

beam-hardening artifacts
20
correction function
20
severe beam-hardening
16
projection data
16
conventional methods
12
proposed method
12
beam-hardening
11
correction
9
function prior-image-based
8
dark bands
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!