A simple and economical method has been developed for simultaneous determination of human serum 25-hydroxyvitamin D (25OHD) and 25-hydroxyvitamin D (25OHD) using Ultra Performance Liquid Chromatography (UPLC). Non-human matrix of 4% BSA was used to construct the calibration curve and in quality control samples' preparation to avoid interference of the endogenous 25-hydroxyvitamin D (25OHD) present in the human serum. 25OHD, 25OHD and dodecanophenone (internal standard, IS) were separated on a CORTECS solid-core particle column and monitored by photodiode array detector at wavelength of 265 nm within five min run time. The relationship between 25OHD concentration and peak area ratio (25OHD:IS) was linear over the range of 12.5 - 200 nM with mean correlation coefficients (r) >0.998. The limit of detection (LOD) for 25OHD and 25OHD was 3.00 nM and 3.79 nM, while the lower limit of quantification (LLOQ) was 9.11 nM and 11.48 nM, respectively. High repeatability was obtained for both isomers with intra-day CV% <5.6% and <5.3% for inter-day assay. This method was further tested with a commercial lyophilized serum control with an accuracy of 92.87-108.31% and applied on 214 human serum samples. In summary, this validated method with BSA can be reliably applied for routine quantification of 25OHD in adults.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.cca.2018.06.024DOI Listing

Publication Analysis

Top Keywords

human serum
12
25-hydroxyvitamin 25ohd
12
simultaneous determination
8
ultra performance
8
performance liquid
8
liquid chromatography
8
25ohd
8
25ohd 25ohd
8
25-hydroxyvitamin
5
determination 25-hydroxyvitamin
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!