Purpose: Intraoperative radiotherapy using The INTRABEAM System (Carl Zeiss Meditec AG, Jena, Germany), a miniature low-energy x-ray source, has proven to be an effective modality in the treatment of breast cancer. However, some uncertainties remain in its dosimetry. In this work, we investigated the INTRABEAM system dosimetry by performing ionization chamber and radiochromic film measurements of absorbed dose in a water phantom.

Methods: Ionization chamber measurements were performed with a PTW 34013 parallel-plate soft x-ray chamber at source to detector distances of 5 to 30 mm calculated using (a) the dose formula consistent with the TARGIT breast protocol (TARGIT), (b) the formula recommended by the manufacturer (Zeiss), and (c) the recently proposed CQ formalism of Watson et al. (Physics in Medicine & Biology, 2018;63:015016) EBT3 Gafchromic film measurements were made at the same depths in water. To account for the energy dependence of EBT3 film, multiple dose response calibration curves were employed across a range of photon beam qualities relevant to the INTRABEAM spectrum in water.

Results: At all depths investigated, the TARGIT dose was significantly lower than that measured by the Zeiss and CQ methods, as well as film. These dose differences ranged from 14% to as large as 80%. In general, the doses measured by film, and the Zeiss and CQ methods were in good agreement to within measurement uncertainties (5-6%).

Conclusions: These results suggest that the TARGIT dose underestimates the physical dose to water from the INTRABEAM source. Understanding the correlation between the TARGIT and physical dose is important for any studies wishing to make dosimetric comparisons between the INTRABEAM and other radiation emitting devices.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.13059DOI Listing

Publication Analysis

Top Keywords

ionization chamber
12
film measurements
12
x-ray source
8
chamber radiochromic
8
radiochromic film
8
intrabeam system
8
dose
8
dose water
8
targit dose
8
zeiss methods
8

Similar Publications

Purpose: This study aims to compare treatment plans created using RapidPlan and PlanIQ for twelve patients with prostate cancer, focusing on dose uniformity, dose reduction to organs at risk (OARs), plan complexity, and dose verification accuracy. The goal is to identify the tool that demonstrates superior performance in achieving uniform target dose distribution and reducing OAR dose, while ensuring accurate dose verification.

Methods: Dose uniformity in the planning target volume, excluding the rectum, and dose reduction in the OARs (the rectum and bladder) were assessed.

View Article and Find Full Text PDF

Establishing In-vivo brain microdialysis for comparing concentrations of a variety of cortical neurotransmitters in the awake rhesus macaque between different cognitive states.

J Neurosci Methods

January 2025

Cognitive Neuroscience Laboratory, German Primate Center - Leibniz Institute for Primate Research, Goettingen, Germany; Faculty of Biology and Psychology, University of Goettingen, 37077 Goettingen, Germany.

Background: Neuronal activity is modulated by behavior and cognitive processes. The combination of several neurotransmitter systems, acting directly or indirectly on specific populations of neurons, underlie such modulations. Most studies with non-human primates (NHPs) fail to capture this complexity, partly due to the lack of adequate methods for reliably and simultaneously measuring a broad spectrum of neurotransmitters while the animal engages in behavioral tasks.

View Article and Find Full Text PDF

Assessment of the Performance of the Dose Calibrator Used in Radioactivity Measurement.

Indian J Nucl Med

November 2024

Center for Research and Production of Radioisotopes, Dalat Nuclear Research Institute, Vietnam Atomic Energy Institute (VINATOM), Da Lat City, Lam Dong Province, Vietnam.

Aims: This study aimed to evaluate the principal technical characteristics of a well-type gas-filled ionization chamber dose calibrator used in measuring radiopharmaceutical activity, namely accuracy, repeatability, and linearity. Furthermore, this work also explored the correlation between the device's response and the position and volume of the radiopharmaceutical I-131.

Materials And Methods: Experimental measurements were conducted on the ATOMLAB 500 dose calibrator using NIST traceable Cs-137 source to determine the accuracy and repeatability.

View Article and Find Full Text PDF

Challenges in extracting and characterizing electrolytes from automotive lithium-ion cells.

Anal Chim Acta

January 2025

University Regensburg, Institute of Analytical Chemistry, Universitätsstrasse 31, 93053, Regensburg, Germany. Electronic address:

Background: The demand for lithium-ion cells in the automotive industry is rapidly growing due to the increasing electrification of the transportation sector. The electrolyte composition plays a critical role in determining the lifetime and performance of these large-format cells. Additionally, advancements in this field are leading to frequent changes in both electrode materials and electrolyte formulations.

View Article and Find Full Text PDF

Purpose: The self-shielding radiosurgery system ZAP-X consists of a 3 MV linear accelerator and eight round collimators. For this system, it is a common practice to perform the reference dosimetry using the largest 25 mm diameter collimator at a source-to-axis distance (SAD) of 45 cm with the PTW Semiflex3D chamber placed at a measurement depth of 7 mm in water. Existing dosimetry protocols do not provide correction for these measurement conditions.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!