Objectives: To set up ELISA for detection of atrazine with high precision.

Methods: The reaction condition of indirect-ELISA was optimized, including atrazine-ovalbumin(AT-OVA) concentration and primary antibody concentration, organic solvent, goat anti-rat immunoglobin G-horseradish peroxidase(IgG-HRP) concentration. The actual samples were detected by the ELISA method established in our laboratory. Then the ELISA method was compared with the HPLC.

Results: The specification curve of indirect-ELISA was set up after optimization. The relation coefficient R=0.9958. The limit of detection (LOD) was 1.972 ng/ml. The percent recovery of the actual samples was in range of 80%~120%. The ELISA detection sensitivity was higher than the HPLC in the range of 0 ng/ml~6 ng/ml atrazine.

Conclusions: The ELISA to detect atrazine has good specificity and high precision. And it can be applied in testing real atrazine samples replacing of the large-scale instrument.