We present a fiber-optic dual-confocal laser caliper method for noncontact high-precision sensing and measuring thickness and refractive index of intraocular lens (IOL) implants. The principle of the method is based on sensing and measuring the confocal intensity response of the laser beam reflection from the opposite object surfaces, which provides the advanced feature of having no limitations on the object shape, thickness, and transparency. Using single-mode optical fibers and a 658-nm laser source, the thickness measurement accuracy was assessed to be as high as 5 μm. In addition, refractive index of a transparent object with thickness smaller than the working distance of the focusing lenses can be measured. The thickness and refractive index of a planoconvex IOL were measured with a high accuracy.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1117/1.JBO.23.6.067004 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!