Objective: Cleared blood glucose monitors (BGMs) for personal use may not always deliver levels of accuracy currently specified by international and U.S. regulatory bodies. This study's objective was to assess the accuracy of 18 such systems cleared by the U.S. Food and Drug Administration representing approximately 90% of commercially available systems used from 2013 to 2015.
Research Design And Methods: A total of 1,035 subjects were recruited to have a capillary blood glucose (BG) level measured on six different systems and a reference capillary sample prepared for plasma testing at a reference laboratory. Products were obtained from consumer outlets and tested in three triple-blinded studies. Each of the three participating clinical sites tested a different set of six systems for each of the three studies in a round-robin. In each study, on average, a BGM was tested on 115 subjects. A compliant BG result was defined as within 15% of a reference plasma value (for BG ≥100 mg/dL [5.55 mmol/L]) or within 15 mg/dL (0.83 mmol/L) (for BG <100 mg/dL [5.55 mmol/L]). The proportion of compliant readings in each study was compared against a predetermined accuracy standard similar to, but more lenient than, current regulatory standards. Other metrics of accuracy included the overall compliance proportion; the proportion of extreme outlier readings differing from the reference value by >20%; modified Bland-Altman analysis including average bias, coefficient of variation, and 95% limits of agreement; and proportion of readings with no clinical risk as determined by the Surveillance Error Grid.
Results: The different accuracy metrics produced almost identical BGM rankings. Six of the 18 systems met the predetermined accuracy standard in all three studies, 5 systems met it in two studies, and 3 met it in one study. Four BGMs did not meet the accuracy standard in any of the three studies.
Conclusions: Cleared BGMs do not always meet the level of analytical accuracy currently required for regulatory clearance. This information could assist patients, professionals, and payers in choosing products and regulators in evaluating postclearance performance.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.2337/dc17-1960 | DOI Listing |
J Adv Nurs
January 2025
Key Laboratory of Birth Defects and Related Diseases of Women and Children (Sichuan University), Ministry of Education, Chengdu, China.
Aim: To identify the barriers and enablers in the implementation of evidence-based physical activity (PA) programmes for the improvement of health outcomes among pregnant women with gestational diabetes mellitus (GDM), and to develop strategies for implementing this evidence in clinical practice.
Methods: A convergent mixed-methods study was conducted, integrating a descriptive qualitative research design with a cross-sectional survey. In-depth interview was used to collect the views and cognitions about physical activity from medical staff, leaders and pregnant women.
Naunyn Schmiedebergs Arch Pharmacol
January 2025
Department of Pharmaceutics, National Institute of Pharmaceutical Education and Research, Raebareli (NIPER-R), Transit Campus, Bijnor-Sisendi Road, Sarojini Nagar, Near CRPF Base Camp, Lucknow, UP, Lucknow, 226002, India.
Atopic dermatitis (AD) is a chronic skin inflammatory ailment commonly observed in young children and adults. Various therapeutic modalities are already explored for mitigation of AD but for prolong application very few modalities are recommended. Considering these challenges, we have successfully developed gliclazide-loaded hydrogels using the physical dispersion method.
View Article and Find Full Text PDFMetab Brain Dis
January 2025
Department of Neurology, Beijing Tiantan Hospital, Capital Medical University, Beijing, China.
In the acute stage of stroke, stress hyperglycemia is common in both diabetic and nondiabetic patients. The associations between stress hyperglycemia and functional outcomes, as well as stroke recurrence were heterogeneous in previous studies. We aimed to demonstrate these associations in a general population of patients with ischemic stroke and transient ischemic attack (TIA).
View Article and Find Full Text PDFEur J Haematol
January 2025
Faculty of Medicine, Tel Aviv University, Tel Aviv-Yafo, Israel.
Background: Bone marrow examination (BME) is the gold standard of diagnosing myelodysplastic syndromes (MDS).
Problems: it is invasive, painful, causing possible bleeding, inaccurate (aspirate hemodilution), and subjective (inter-observer interpretation discordance). We developed non-invasive diagnostic tools: A logistic regression formula [LeukRes 2018], then a web algorithm using 10 variables (age, gender, Hb, MCV, WBC, ANC, monocytes, PLT, glucose, creatinine) to diagnose/exclude MDS [BldAdv 2021].
Alzheimers Res Ther
January 2025
Functional Imaging Unit, Department of Clinical Physiology and Nuclear Medicine, Copenhagen University Hospital Rigshospitalet, Valdemar Hansens Vej 1-23, Glostrup, 2600, Denmark.
Background: Accumulation of β-amyloid (Aβ) in the brain is a hallmark of Alzheimer's Disease (AD). Cerebral deposition of Aβ initiates deteriorating pathways which eventually can lead to AD. However, the exact mechanisms are not known.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!