Background: It remains undetermined whether second-generation drug-eluting stents (G2-DESs) outperform first-generation DESs (G1-DESs) in patients with acute coronary syndrome (ACS). We aimed to compare the efficacy and safety of G1-DES and G2-DES in ACS patients in a high-volume cardiovascular center.

Methods: In 2013, 10,724 consecutive patients underwent percutaneous coronary intervention in our institution. We included 4037 patients with ACS who underwent exclusively G1-DES or G2-DES implantation (n = 364 and n = 3673, respectively). We used propensity score matching to minimize the imbalance between the G1-DES and G2-DES groups and followed patients for 2 years. The efficacy endpoints were major adverse cardiac events (MACEs) and its components including target vessel-related myocardial infarction (TV-MI), target vessel revascularization/target lesion revascularization (TVR/TLR), and cardiac death. The safety endpoint was stent thrombosis. Continuous variables were compared by Mann-Whitney U-test, and categorical variables were compared using Pearson's Chi-square or Fisher's exact test. Kaplan-Meier curves were constructed to compare the event-free survival rates, and multivariate Cox proportional hazards regression analysis was used to assess whether stent type was an independent risk factor for the efficacy and safety endpoints.

Results: At the 2-year follow-up, the results for MACE and it components, as well as stent thrombosis, were similar for G1-DES and G2-DES (MACE, 5.2% vs. 4.3%, χ = 0.514, P = 0.474; TV-MI, 0.8% vs. 0.4%, P = 0.407; TVR, 4.9% vs. 3.7%, χ = 0.939, P = 0.333; TLR, 3.8% vs. 2.5%, χ = 1.610, P = 0.205; cardiac death, 0.3% vs. 0.5%, P = 0.670; and stent thrombosis, 0.5% vs. 0.4%, P > 0.999). Kaplan-Meier analysis indicated similar event-free survival rates between G1-DES and G2-DES after propensity score matching (all: log-rank P > 0.05). Multivariate analysis demonstrated that stent type was not an independent risk factor for the efficacy and safety endpoints (MACE, hazard ratio [HR] = 0.805, 95% confidence interval [CI]: 0.455-1.424, P = 0.456; TV-MI, HR = 0.500, 95% CI: 0.101-2.475, P = 0.395; TVR, HR = 0.732, 95% CI: 0.403-1.330, P = 0.306; TLR, HR = 0.629, 95% CI: 0.313-1.264, P = 0.193; cardiac death, HR = 1.991, 95% CI: 0.223-17.814, P = 0.538; and stent thrombosis, HR = 0.746, 95% CI: 0.125-4.467, P = 0.749).

Conclusion: G1-DES and G2-DES have similar efficacy and safety profiles in ACS patients at the 2-year follow-up.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6006822PMC
http://dx.doi.org/10.4103/0366-6999.233959DOI Listing

Publication Analysis

Top Keywords

g1-des g2-des
24
efficacy safety
20
stent thrombosis
16
cardiac death
12
second-generation drug-eluting
8
drug-eluting stents
8
patients acute
8
acute coronary
8
coronary syndrome
8
acs patients
8

Similar Publications

Background: There are limited data on the long-term stent-related adverse events as related to the duration of dual antiplatelet therapy (DAPT) in second-generation (G2) drug-eluting stents (DES) compared with first-generation (G1) DES.

Objectives: This study sought to compare the long-term stent-related outcomes of G2-DES with those of G1-DES.

Methods: The study group consisted of 15,009 patients who underwent their first coronary revascularization with DES from the CREDO-Kyoto PCI/CABG (Coronary Revascularization Demonstrating Outcome Study in Kyoto Percutaneous Coronary Intervention/Coronary Artery Bypass Grafting) Registry Cohort-2 (first-generation drug-eluting stent [G1-DES] period; n = 5,382) and Cohort-3 (second-generation drug eluting stent [G2-DES] period; n = 9,627).

View Article and Find Full Text PDF

Over the past 40 years, the safety and efficacy of percutaneous coronary intervention has dramatically improved by overcoming several challenges. The introduction of drug-eluting stent (DES) in particular was a major breakthrough in interventional cardiology. Compared to bare-metal stents, first-generation DES (G1-DES) has dramatically reduced the rates of in-stent restenosis and subsequent target lesion revascularization.

View Article and Find Full Text PDF

Comparison of Efficacy and Safety between First- and Second-Generation Drug-Eluting Stents in Patients with Acute Coronary Syndrome.

Chin Med J (Engl)

June 2018

Department of Cardiology, State Key Laboratory of Cardiovascular Disease, Fuwai Hospital, National Center for Cardiovascular Diseases, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100037, China.

Background: It remains undetermined whether second-generation drug-eluting stents (G2-DESs) outperform first-generation DESs (G1-DESs) in patients with acute coronary syndrome (ACS). We aimed to compare the efficacy and safety of G1-DES and G2-DES in ACS patients in a high-volume cardiovascular center.

Methods: In 2013, 10,724 consecutive patients underwent percutaneous coronary intervention in our institution.

View Article and Find Full Text PDF

Background: The comparative efficacy of second-generation (G2) vs. first-generation (G1) drug-eluting stents (DES) for calcified coronary lesions is unknown.

Methods and results: We compared the 3-year clinical outcomes of patients with G1- or G2-DES according to the presence or absence of calcified coronary lesions as assessed in an angiographic core laboratory using data from 2 large-scale prospective multicenter randomized trials, RESET and NEXT.

View Article and Find Full Text PDF

Background: Lots of trials demonstrate that second-generation drug-eluting stents (G2-DES), with their improved properties, offer significantly superior efficacy and safety profiles compared to first generation DES (G1-DES) for patients with coronary artery disease (CAD) receiving percutaneous coronary intervention (PCI). This study aimed to verify the advantage of G2-DES over G1-DES in Chinese patients with stable CAD (SCAD).

Methods: For this retrospective observational analysis, 2709 SCAD patients with either G1-DES (n = 863) or G2-DES (n = 1846) were enrolled consecutively throughout 2013.