Assessment of straylight and the modulation transfer function of intraocular lenses with centrally localized opacification associated with the intraocular injection of gas.

J Cataract Refract Surg

From the David J. Apple Center for Vision Research, Department of Ophthalmology (Łabuz, Yildirim, Khoramnia, Auffarth), University Hospital Heidelberg, Heidelberg, Germany; the Netherlands Institute for Neuroscience (van den Berg), Royal Netherlands Academy of Arts and Sciences, Amsterdam, the Netherlands.

Published: May 2018

Purpose: To assess the optical quality of intraocular lenses (IOLs) explanted because of opacification after the intraocular injection of gas.

Setting: David J. Apple Laboratory, Heidelberg, Germany.

Design: Experimental study.

Methods: Four hydrophilic acrylic IOLs were studied, each with a centrally localized round opacification pattern associated with the intraocular use of gas. Laboratory analysis included gross examination with a light microscope, followed by alizarin and von Kossa staining. Optical quality was assessed by examining the modulation transfer function (MTF) and straylight. Results were compared with those of a control IOL and normative data for straylight of the crystalline lens. The following parameters were derived from image analysis: opacified surface fraction, light loss in the opacified surface, and the area and number of granules. The relationship between straylight increase and those parameters was studied.

Results: Fine granules were identified on the IOL surface and subsurface. The granules stained positive for calcium, and the MTF levels of 2 IOLs dropped markedly. The other 2 showed relatively minor changes. The straylight was extremely increased in 3 IOLs up to (and above) a level of that of a cataractous lens. A proportional relationship was found between straylight and the morphological parameters from image analysis.

Conclusions: Intraocular lenses with centrally localized opacification have a strong potential for deteriorating optical performance. However, the optical quality might differ depending on the morphology of opacification. A serious straylight increase was found in most of these IOLs, suggesting that affected patients may suffer from glare-related symptoms.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jcrs.2018.01.033DOI Listing

Publication Analysis

Top Keywords

intraocular lenses
12
centrally localized
12
optical quality
12
modulation transfer
8
transfer function
8
lenses centrally
8
localized opacification
8
associated intraocular
8
intraocular injection
8
opacified surface
8