Introduction: The purpose of this study was to compare acetabular cup positioning using an imageless navigation-assisted total hip arthroplasty (THA) to a case-matched control group using the conventional technique in Asian physique.
Methods: Case matching controlled study using retrospectively collected data for THAs was done. Matching criteria were gender, age ± 5 years, and body mass index ± 5 kg/m. Thirty patients who had navigation THAs were manually matched to a conventional group of 30 patients. After measuring anteversion and inclination on 3-D CT, final cup anteversion was calculated using the mathematical formula (anteversion = arctan [tan (anatomic anteversion) × sin (inclination)]).
Results: No significant difference was observed between mean inclination ( p = 0.420), anteversion ( p = 0.697), and the mean deviation from 15° for anteversion ( p = 0.649). However, there was a statistical significant difference for mean deviation from 40° for inclination ( p = 0.027). There was a significant difference in the percentage of correctly placed acetabular cup in safe zones with inclination ( p = 0.039).
Conclusions: Although soft tissue overlying the pubic tubercle interfered with registration of the bony landmark and thereby induced inaccuracy of anteversion angles in the navigation group, the use of imageless navigation-assisted technique increases the placement within the safe zone for inclination and enhances the accuracy and the precision of inclination of the acetabular cup relative to the conventional THA rather than anteversion. This finding may be due to the effect of the patient's physique on the anterior pelvic plane (APP)-based reference system of imageless navigation.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1177/2309499018780755 | DOI Listing |
Hip Int
January 2025
Orthopaedic Clinic Lucerne, Lucerne, Switzerland.
Introduction: The RM Pressfit vitamys is an uncemented, titanium particle-coated, isoelastic monoblock cup made of vitamin E blended highly cross-linked polyethylene. We addressed the following questions: (1) What are the clinical and (2) radiographic outcomes 10 years after implantation? (3) What is the revision rate?
Methods: In this prospective observational study in a tertiary care centre we investigated all consecutive cases of total hip replacement with the RM Pressfit vitamys cup between September 2009 and November 2011. It was implanted in 162 hips, 49.
Arthroplasty
January 2025
Sant'Andrea Hospital, Faculty of Medicine and Psychology, Sapienza University of Rome, 00162, Rome, Italy.
Background: Proper positioning of a total hip arthroplasty (THA) plays a crucial role in the success and long-term survivorship of the implant. Cup positioning within the Lewinnek Safe Zone (LSZ) does not, however, avoid implant dislocation. Thus, the concept of a functional cup position has been introduced.
View Article and Find Full Text PDFExpert Rev Med Devices
January 2025
Department of Orthopaedic Surgery, Kansai Medical University, Hirakata, Osaka, Japan.
Introduction: There is a worldwide trend toward 'revisiting' cemented total hip arthroplasty (THA). In Japan, however, cemented THAs accounts for 11%, and the percentage of cemented hemiarthroplasty is estimated to be less than 10%. This review was designed to reconsider the option of cemented THA and to encourage policy changes in Japan to support the best possible care for patients.
View Article and Find Full Text PDFBone Jt Open
January 2025
Department of Orthopaedic Surgery, Ehime University Graduate School of Medicine, Toon, Japan.
Aims: Excellent outcomes have been reported following CT-based robotic arm-assisted total hip arthroplasty (rTHA) compared with manual THA; however, its superiority over CT-based navigation THA (nTHA) remains unclear. This study aimed to determine whether a CT-based robotic arm-assisted system helps surgeons perform accurate cup placement, minimizes leg length, and offsets discrepancies more than a CT-based navigation system.
Methods: We studied 60 hips from 54 patients who underwent rTHA between April 2021 and August 2023, and 45 hips from 44 patients who underwent nTHA between January 2020 and March 2021 with the same target cup orientation at the Department of Orthopedic Surgery at Ozu Memorial Hospital, Japan.
Indian J Orthop
January 2025
Trauma & Orthopaedics, Hinchingbrooke Hospital, North-West Anglia NHS Foundation Trust, Huntingdon, UK.
Background: Obesity has been consistently proven to be associated with an increased risk of dislocation following total hip arthroplasty (THA). As the prevalence of obesity continues to rise globally, it is of vital importance to minimise risks, including dislocation rates, in these patients undergoing THA.
Methods: We describe a series of patients with obesity, morbid obesity (BMI ≥ 40) and super-obesity (BMI ≥ 50) undergoing THA at our institution over a 10-year period using a dual-mobility acetabular cup.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!