Introduction: Since 1996 the active middle-ear implant Vibrant Soundbridge (VSB) is used to treat mild-to-severe sensorineural hearing losses. The former standard surgical approach for incus vibroplasty included a mastoidectomy and a posterior tympanotomy, crimping the Floating-Mass Transducer (FMT) to the long incus process. The introduction of the short process (SP) coupler allows the attachment of the FMT to the incus body making the procedure of a posterior tympanotomy unnecessary. The aim of this study was to evaluate the duration of surgery (surgical case length) and to compare the audiological results of SP coupling with the former standard surgical approach to the long incus process.

Materials And Methods: Patients undergoing an incus vibroplasty between October 2009 and February 2016 were included in this cohort study. The patients received an incus vibroplasty with crimping the FMT to the long incus process (former standard application) (n = 25) and with the SP coupler (n = 17) respectively. The surgical case length as well as the functional audiological outcome was assessed 12 months postoperatively using pure-tone audiometry and speech audiometry.

Results: The surgical case length was significantly shorter in the SP coupler group compared to the standard application incus vibroplasty (85 ± 29 min vs. 114 ± 50 min). Additionally, patients receiving the SP coupler had an increased speech perception if compared to the standard application (Freiburger monosyllables at 65 dB SPL: 76.1 ± 16.1% vs. 66.2 ± 23.5%).

Conclusion: The SP coupler leads to a shortened time of surgery and by the less invasive surgery to reduced risks for facial nerve and chorda tympani. Speech perception is significantly improved by SP coupling compared to classic long incus coupling.

Download full-text PDF

Source
http://dx.doi.org/10.1097/MAO.0000000000001830DOI Listing

Publication Analysis

Top Keywords

incus vibroplasty
16
long incus
16
incus process
12
surgical case
12
case length
12
standard application
12
incus
10
process introduction
8
standard surgical
8
surgical approach
8

Similar Publications

Background: The "floating mass transducer" (FMT) of the Vibrant Soundbridge was initially attached to the long process of the incus (LP). In patients missing an incus, coupling to the stapes head using the Vibroplasty CliP-coupler is used. Its main limitation is the height, leading to potential contact with the tympanic membrane.

View Article and Find Full Text PDF

Objectives: The active middle ear implant, Vibrant Soundbridge (VSB), can be implanted with a variety of couplers. Hearing outcome after implantation has been investigated in both temporal bone (TB) experiments and patient studies, but the relationship between experimental and clinical data is still weak in the literature. Therefore, experimental data from TB experiments should be compared with patient data in a retrospective study, in which the floating mass transducer is used with couplers of the third generation.

View Article and Find Full Text PDF

Less than 20% of patients with sensorineural hearing loss (HL) provided with the Vibrant Soundbridge (VSB) experience a progressive HL and warrant cochlear implantation (CI). The aim of this study was to identify possible predictors of progressive HL prior to VSB implantation. This retrospective study included all consecutive ears with sensorineural HL provided with the VSB between 1998 and 2016.

View Article and Find Full Text PDF

[Historical overview and surgical applications of the Vibrant Soundbridge active middle ear implant].

Orv Hetil

October 2021

1 Pécsi Tudományegyetem, Általános Orvostudományi Kar, Fül-Orr-Gégészeti és Fej-Nyaksebészeti Klinika, Pécs, Munkácsy M. u. 2., 7621.

Összefoglaló. A Vibrant Soundbridge aktív középfül-implantátum vezetéses, sensorineuralis és kevert típusú halláscsökkenés esetén is megoldást jelenthet a beteg hallásrehabilitációja során. Sensorineuralis halláscsökkenés esetén akkor indikálható, ha a légvezetéses hallásküszöb nem haladja meg a 80-85 dBHL-t a 3-4 kHz frekvenciákon, vezetéses, illetve kevert típusú halláscsökkenés esetén pedig akkor, ha a csontvezetéses hallásküszöb nem haladja meg a 45-65 dBHL-t a 0,5-4 kHz közti frekvenciatartományban.

View Article and Find Full Text PDF

Objective: The study shows the long-term effectiveness, safety, and quality of life after Vibrant Soundbridge (VSB) implantation in sensorineural hearing loss (SNHL) using the short process coupler (SP) or the long process coupler (LP).

Methods: This retrospective study evaluated 77 VSB cases. Follow-up (F/U) time-dependent objective measurements (audiological outcomes), subjective data collection (quality-of-life questionnaire), and safety measures are presented.

View Article and Find Full Text PDF