Background: Contrast-enhanced cardiac computed tomography (CE-CT) provides useful substrate characterization in patients with ventricular tachycardia (VT).
Objective: The purpose of this study was to describe the association between endocardial electrogram measurements and myocardial characteristics on CE-CT, in particular the field of view of electrogram features.
Methods: Fifteen patients with postinfarct VT who underwent catheter ablation with preprocedural CE-CT were included. Electroanatomic maps were registered to CE-CT, and myocardial attenuation surrounding each endocardial point was measured at a radius of 5, 10, and 15 mm. The association between endocardial voltage and attenuation was assessed using a multilevel random effects linear regression model, clustered by patient, with best model fit defined by highest log likelihood.
Results: A total of 4698 points were included. There was a significant association of bipolar and unipolar voltage with myocardial attenuation at all radii. For unipolar voltage, the best model fit was at an analysis radius of 15 mm regardless of the mapping catheter used. For bipolar voltage, the best model fit was at an analysis radius of 15 mm for points acquired with a conventional ablation catheter. In contrast, the best model fit for points acquired with a multipolar mapping catheter was at an analysis radius of 5 mm.
Conclusion: Myocardial attenuation on CE-CT indicates a smaller myocardial field of view of bipolar electrograms using multipolar catheters with smaller electrodes in comparison to standard ablation catheters despite similar interelectrode spacing. Smaller electrodes may provide improved spatial resolution for the definition of myocardial substrate for VT ablation.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8262627 | PMC |
http://dx.doi.org/10.1016/j.hrthm.2018.05.031 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!