Background: Postoperative squeaking in patients who applied the fourth-generation ceramic bearing in primary hip replacement has not been reported systematically; we aim to study the squeaking incidence in the fourth-generation ceramic bearing and related risk factors for squeaking, and we also attempt to explore the relationship between squeaking and prosthetic brands.

Methods: The PubMed, Embase, and Cochrane library were searched, and 14 articles were finally included. Patients' demographic data, surgical-related information, and prosthesis data were extracted. The occurrence rate of squeaking was calculated by meta-analysis, and subgroup analysis was performed based on prosthetic brands and follow-up time. Regression analysis was further applied to investigate the relationship between various risk factors and squeaking.

Results: The squeaking incidence in patients with the fourth-generation ceramic bearing was 3%. Age, gender, body mass index, and abduction and anteversion angles of acetabular cup might have no influence on squeaking. The squeaking incidence was significantly high with the presence of Delta Motion cup (DePuy, Warsaw, Indiana) and Secure-Fit stem (Stryker, Kalamazoo, MI), and the overall incidence of DePuy femoral stem was relatively small except for the Summit femoral stem. And there was no significant difference of squeaking incidence between less than 5-year and more than or equal to 5-year follow-up subgroups.

Conclusions: In our study, squeaking in the fourth-generation ceramic bearing occurred at a rate of 3%; occurrence rate was high when the Delta Motion cup was applied. We hope for more relevant researches to focus on this issue.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5984797PMC
http://dx.doi.org/10.1186/s13018-018-0841-yDOI Listing

Publication Analysis

Top Keywords

fourth-generation ceramic
16
ceramic bearing
16
squeaking incidence
16
squeaking
11
squeaking fourth-generation
8
hip replacement
8
study squeaking
8
risk factors
8
occurrence rate
8
delta motion
8

Similar Publications

Article Synopsis
  • * Out of the reviewed surgeries, there were only 33 revisions (0.36%) due to CoC bearing fractures, primarily involving the third-generation components, with most fractures occurring in 28 mm ceramic heads.
  • * The research indicates a low overall risk of CoC bearing fractures, with younger age and being male associated with a higher likelihood of fractures, and highlights the advancements in fourth-generation ceramics in reducing head fracture occurrences.
View Article and Find Full Text PDF

Purpose: To assess utility, benefits, and risks of 4th-generation alumina-zirconia ceramic pairings in elective total hip arthroplasty (THA).

Methods: A comprehensive mixed-methods best-evidence synthesis using data from systematic reviews, randomized controlled trials (RCTs), prospective and retrospective cohort studies, as well as joint replacement registries, was conducted to estimate overall revision and survival rates, periprosthetic infection, bearing fractures, and noise phenomena with 4th-generation alumina-zirconia ceramic versus other tribological couplings in elective THA. The systematic review part across multiple databases was registered with PROSPERO (CRD42023418076), and individual study data were extracted for statistical re-analysis.

View Article and Find Full Text PDF

Background: The translucency of different zirconia generations at each time point after thermocycling aging is still lacking.

Methods: Four zirconia materials were used with a total of 60 samples produced from monolithic third generation (5Y) 5 mol% yttria-stabilized zirconia polycrystalline ceramic and fourth generation zirconia (4Y) 4 mol% yttria-stabilized zirconia polycrystalline ceramic, represented by [group1:[CM-5Y] Ceramill Zolid fx (3rd generation zirconia) (Amann Girrbach, Koblach, Austria), group 2:[CM-4Y] Ceramill Zolid HT + (4th generation zirconia) (Amann Girrbach, Koblach, Austria), group 3:[CC-5Y] Cercon XT/ML (Dentsply Sirona, Germany) (3rd generation), and group 4:[CC-4Y] Cercon HT/ML (Dentsply Sirona, Germany) (4th generation)]. The L*a*b* figures were measured by using a spectrophotometer at baseline and after 10,000, 30,000, and 50,000 cycles of thermocycling.

View Article and Find Full Text PDF

Background: We previously reported the mean 6-year outcomes of total hip arthroplasty (THA) using fourth-generation ceramic-on-ceramic (CoC) articulations. With extended observations, this study aimed to report the mean 12-year outcomes focusing on bearing-related complications of CoC THA.

Methods: We reviewed primary CoC THAs performed between May 2009 and April 2012 at a single institution.

View Article and Find Full Text PDF

Zirconium carbide (ZrC) ceramics have a high melting point, low neutron absorption cross section, and excellent resistance to the impact of fission products and are considered to be one of the best candidate materials for fourth-generation nuclear energy systems. ZrC ceramics with a high relative density of 99.1% were successfully prepared via pressureless sintering using a small amount of MoSi as an additive.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!