Background: To review the procedural safety and postimplantation complications of Watchman device implanted at 2 community hospitals for primary prevention of systemic embolization in patients with nonvalvular atrial fibrillation (NVAF) who were not candidates for long-term oral anticoagulation (OAC).

Methods: This was a retrospective case series of 48 patients carried out in 2 community hospitals in the United States. Patients with NVAF who had a CHADS2 higher than 2 or CHADS2VASc2 (congestive heart failure, hypertension, age ≥75 years, diabetes mellitus, prior stroke or transient ischemic attack [TIA] or thromboembolism, vascular disease, age 65-74 years, and female gender) score of 3 or higher and were not candidates for long-term OAC. These patients were selected for implantation of Watchman device. They were followed up at 45 days, 6 months, 9 months, and 12 months after implantation of Watchman device to assess for complications involving the device and to determine if anticoagulation could be discontinued at the 45 days follow-up. They were monitored for any systemic thromboembolism while off anticoagulation.

Results: The success rate of device implantation was 98% (48 of 49). Only a single patient could not get Watchman implantation because of unfavorable left atrial appendage anatomy. Access-related and device implantation-related complications were zero (0%). At 45 days follow-up and end of follow-up duration, the rate of thrombus formation on the Watchman device was 4% (2 of 48). One patient had TIA after warfarin discontinuation.

Conclusion: With improved procedural technique and well-trained operators, Watchman implantation is feasible in a community hospital also.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jstrokecerebrovasdis.2018.05.016DOI Listing

Publication Analysis

Top Keywords

watchman device
16
community hospital
8
united states
8
left atrial
8
atrial appendage
8
device
8
community hospitals
8
candidates long-term
8
implantation watchman
8
months months
8

Similar Publications

Pseudo Peridevice Leak After LAA Closure Device Implantation Unmasking a Persistent Left Superior Vena Cava.

JACC Clin Electrophysiol

November 2024

Department of Cardiovascular Medicine, Mayo Clinic, Rochester, Minnesota, USA; Department of Cardiovascular Medicine, Mayo Clinic, LaCrosse, Wisconsin, USA. Electronic address:

View Article and Find Full Text PDF

Background: Research on the impact of angiographically detected residual trabeculation after left atrial appendage closure (LAAC) is limited.

Objectives: To investigate the incidence, characteristics, and clinical implications of angiographically detected residual trabeculation after LAAC using the WATCHMAN device.

Methods: We analyzed 1350 consecutive patients with atrial fibrillation undergoing LAAC using the WATCHMAN device from the OCEAN-LAAC registry, which is a prospective ongoing, multicenter Japanese registry.

View Article and Find Full Text PDF

Outcomes of Concomitant Atrial Fibrillation Ablation and Left Atrial Appendage Closure: A Retrospective Single-Center Experience.

JACC Adv

December 2024

Section of Clinical Cardiac Electrophysiology, Department of Cardiovascular Medicine, Heart, Vascular and Thoracic Institute, Cleveland Clinic, Cleveland, Ohio, USA.

Background: Catheter ablation is an effective therapy in the management of atrial fibrillation (AF). Left atrial appendage closure (LAAC) is an alternative to anticoagulation for stroke prevention in patients with bleeding risks.

Objectives: The purpose of this study was to assess the safety and efficacy of combining AF ablation and LAAC in a single procedure.

View Article and Find Full Text PDF

Patients with atrial fibrillation (AF) often present with concomitant significant mitral regurgitation (MR). Percutaneous left atrial appendage closure (LAAC) is indicated for patients with AF to prevent thromboembolism and reduce the need for long-term anticoagulation. Transcatheter edge-to-edge repair (TEER) is recommended for patients with significant MR.

View Article and Find Full Text PDF

Racial and Ethnic Disparities in the Use and Outcomes With WATCHMAN FLX: A SURPASS Analysis of the NCDR Left Atrial Appendage Occlusion Registry.

J Am Heart Assoc

December 2024

Section of Cardiovascular Medicine Yale University School of Medicine, New Haven, Connecticut and Center for Outcomes Research and Evaluation, Yale-New Haven Hospital New Haven CT USA.

Background: Left atrial appendage occlusion (LAAO) is increasingly used as an alternative to oral anticoagulation for stroke prevention in select patients with atrial fibrillation. Data on outcomes in racial and ethnic minority individuals are limited. This analysis assessed differences in the use and outcomes of LAAO by race and ethnicity in a large national registry.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!