Introduction: Implantable cardioverter-defibrillator (ICD) DF-4 connectors have been introduced to facilitate defibrillator lead connection and to reduce the size of device header. There are limited data regarding the overall performance of those leads and no comparison between different ICD DF-4 leads.

Methods: This is a cohort study of consecutive patients implanted with ICD DF-4 lead system at one University Centre between October 2010 and February 2015. A historical control group of patients with ICD DF-1 lead implantation was used for comparison. The following ICD DF-4 leads were evaluated: St. Jude Medical Durata 7122Q (St. Jude Medical, St. Paul, MN, USA), Medtronic Sprint Quattro Secure 6935 M (Medtronic Inc., Minneapolis, MN, USA), Boston Scientific Endotak Reliance 4-Site 0293 (Boston Scientific, Marlborough, MA, USA), and Boston Scientific Reliance 4-Front 0693. This study evaluated the acute and mid-term performances of those leads as well as complications.

Results: A total of 812 patients (age 63 ± 12 years, 80% male, left ventricular ejection fraction 31 ± 12%) underwent implantation of an ICD DF-4 lead. Acute and follow-up R-wave sensing and threshold were excellent. Compared to implantation, intrinsic R waves were higher at follow-up for Boston Scientific and Medtronic leads, and pacing lead impedances were lower for all leads at first follow-up (P < 0.001). The number of lead dislodgement or failure was similar between all leads. The estimated lead survival rates at 3 years were 95.6% for Boston Scientific Endotak 4-Site, 97.1% for Boston Scientific 4-Front, 97.7% for Medtronic Sprint Quattro, and 97.5% for St. Jude Durata (P  =  0.553).

Conclusion: All ICD DF-4 leads had excellent acute and mid-term electrical performances. Longer follow-up will be necessary to confirm their sustained performance.

Download full-text PDF

Source
http://dx.doi.org/10.1111/pace.13400DOI Listing

Publication Analysis

Top Keywords

icd df-4
20
boston scientific
16
comparison icd
8
df-4 lead
8
jude medical
8
usa boston
8
df-4
6
leads
6
icd
6
lead
5

Similar Publications

The selection of an appropriate implantable cardioverter-defibrillator (ICD) type and implantation strategy involves a myriad of considerations. While transvenous ICDs are standard, the rise of non-transvenous options like subcutaneous ICDs and extravascular ICDs is notable for their lower complication rates. Historical preferences for dual chamber ICDs have shifted to single-chamber ICDs.

View Article and Find Full Text PDF
Article Synopsis
  • Left bundle branch area pacing (LBBAP) shows promise as a safe and effective alternative for cardiac resynchronization therapy (CRT) in patients with implantable cardioverter-defibrillators (ICD).
  • A study assessed the feasibility and safety of a traditional defibrillator lead in the LBBAP location, finding a 62.5% acute procedural success rate with some patients needing lead repositioning.
  • Follow-up results for an average of 3.8 months indicated stable pacing parameters with no inappropriate arrhythmia detection or therapy issues, supporting LBBAP as a viable option.*
View Article and Find Full Text PDF

Initial Experience With Intercostal Insertion of an Extravascular ICD Lead Compatible With Existing Pulse Generators.

Circ Arrhythm Electrophysiol

August 2023

Cardiovascular Department, Sanatorio Italiano, Asunción, Paraguay (A.E.).

Background: This study assessed safety and feasibility of a novel extravascular implantable cardioverter defibrillator (ICD) lead when inserted anteriorly through a rib space and connected to various commercially available ICD pulse generators (PGs) placed in either a left mid-axillary or left pectoral pocket. Currently available or investigational, extravascular-ICDs include a subcutaneous or subxiphoid lead connected to customized extravascular-ICD PGs.

Methods: This novel extravascular-ICD (AtaCor Medical Inc, San Clemente, CA) employs a unique intercostal implant technique and is designed to function with commercial DF-4 ICD PGs.

View Article and Find Full Text PDF

Probe-based qPCR as an alternative to modified Knott's test when screening dogs for heartworm (Dirofilaria immitis) infection in combination with antigen detection tests.

Parasit Vectors

August 2022

Department of Veterinary Pathobiology, College of Veterinary Medicine and Biomedical Sciences, Texas A&M University, 4467 TAMU, College Station, TX, 77843-4467, USA.

Background: Current recommendations for diagnosis of Dirofilaria immitis infection in dogs rely on the detection of antigen produced largely by adult females coupled with the visualization of microfilariae (mf) in the circulation via a microfilaria detection test (MFDT). It is hypothesized that qPCR assays used in parallel with antigen detection tests will perform better in detecting mf than modified Knott's test (MK), when combined with antigen detection. This study compares probe-based qPCR and MK techniques for mf detection used in parallel with the DiroCHEK antigen test to screen for heartworm infection in shelter dogs.

View Article and Find Full Text PDF

Introduction: Data on long-term durability of St Jude Medical Durata defibrillation leads compared to its previous model, the St Jude Medical Riata leads in clinical practice are missing. Aim of the study was to analyze the long-term performance of the Durata defibrillation leads compared to the Riata leads in clinical practice.

Methods And Results: A total of 1407 consecutive patients of a prospective single-centre implantable cardioverter defibrillator (ICD)-registry were analyzed who underwent ICD implantation with a Durata (n = 913) or Riata (n = 494) ICD lead between 2002 and 2017.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!