Unlabelled: Study background: Measurement of delivered pharyngeal pressure during continuous positive airway pressure (CPAP) therapy is not in routine practice due to lack of a simple and affordable technique of intrapharyngeal pressure measurement. To overcome the lack of the gold standard solid-state catheter-tip pressure measurement technology in our set up, we improvised a novel method of pressure measurement and tested its validity in a simulated pharynx.

Methods: A low-cost pressure transducer was improvised by attaching an orogastric tube to its one end. The other end of the orogastric tube was sealed into an artificial pharynx - a 20 ml syringe. The pressure transducer readings were compared with that obtained by a digital manometer attached to the tip of the syringe. Bland-Altman statistic was used to quantify the measurement reliability of the novel method against the digital manometer. Effect of tube length on the measurement agreement was also studied. The developed technique was applied in new-borns.

Results & Conclusion: Pressures measured by this technique were in good agreement with that obtained using a digital manometer. This technique has the potential to be used as an alternative to catheter-tip pressure transducers for bedside pharyngeal pressure measurement in new-born babies, especially in under-resourced setups.

Download full-text PDF

Source

Publication Analysis

Top Keywords

pressure measurement
16
novel method
12
pharyngeal pressure
12
digital manometer
12
pressure
10
catheter-tip pressure
8
pressure transducer
8
orogastric tube
8
measurement
7
description validation
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!